*The technical data are subject to change without prior notice for functional
improvements.
Power Supply
Battery capacity
Storage capacity
Working Temperature
Storage Temperature
Calibration
2 pieces of 1.5V AA size Alkaline batteries
Approx. 3000 tests
500 tests (Storage of serial number, time and date,
measurement result and test number)
-5ºC up to 40ºC
-10ºC up to 50ºC
Every 6 months, see the reminder on the display
when starting the device. ACE can calibrate the
device or the exchangeable sensor, new sensors
are available at ACE Instruments.
For a calibration service you can send the device
to ACE Instruments (see www.ace-instruments.net).
Fixing the hand strap:
1.
Push the button on the back of the device.
2.
Pull the battery cover downwards.
3.
Insert hand strap thorough the small hole in the housing (see picture below).
4.
Fit the strap in the inside grip.
5.
Fix the battery cover on the device.
Technical data*:
Sensor
Accuracy
Detection Range
Warm-up Time
Reaction time
Dimension
Weight
Fuel Cell sensor
+/- 0.050‰ (per mill) with reference to 1.000‰ at 25ºC
(per mill = default setting)
OR +/- 0.005% BAC (Blood Alcohol Concentration)
with reference to 0.100% BAC at 25ºC
OR +/- 0.025mg/l with reference to 0.500mg/l at 25ºC
0.000 - 5.000‰ (per mill = default setting; you can change
the measurement unit in the Administration menu by
choosing “Measurement set.” and then choose “Measure-
ment unit” where you can do the change yourself)
5 seconds up to max. 2 minutes (the higher the
measurement result the longer is the time needed
by the sensor to evaporate the measured alcohol)
Within 10 seconds with reference to 1.000‰ (per mill)
133 x 64 x 33 mm (H x W x D)
198g (battery included)