6. Guidance for Electromagnetic compatibility (EMC)
Details about the electromagnetic compatibility (EMC) of the ACCUNIQ BP800 are given below.
Before using the ACCUNIQ BP800, be sure to read and understand the following information.
1) Guidance and manufacturer’s declaration – electromagnetic emissions
The ACCUNIQ BP800 is intended for use in the electromagnetic environment specified below. The
customer or the user of the ACCUNIQ BP800 should assure that it is used in such an environment.
Emissions test
Compliance
Electromagnetic environment
–
guidance
RF emissions
CISPR 11
Group 1
The ACCUNIQ BP800 uses RF energy only for its internal
function. Therefore, its RF emissions are very low and are
not likely to cause any interference in nearby electronic
equipment.
RF emissions
CISPR 11
Class B
Harmonic emissions
IEC 61000-3-2
Class A
Voltage fluctuations/
flicker emissions
IEC 61000-3-3
Compliance
The ACCUNIQ BP800 is suitable for use in all
establishments, including domestic establishments and
those directly connected to the public low-voltage power
supply network that supplies buildings used for domestic
purposes.
2) Guidance and manufacturer’s declaration – electromagnetic immunity
The ACCUNIQ BP800 is intended for use in the electromagnetic environment specified below. The
customer or the user of the ACCUNIQ BP800 should assure that it is used in such an environment.
Immunity test
IEC 60601
test level
Compliance level
Electromagnetic
environment-
guidance
Electrostatic
discharge(ESD)
IEC 61000-4-2
±6kV: Contact
±8kV: Air
±6kV: Contact
±8kV: Air
Floors should be wood, concrete or
ceramic tile. If floors are covered with
synthetic
material,
the
relative
humidity should be at least 30 %.
Electrical
fast
transition/burst
IEC 61000-4-4
±2kV:
Power
supply lines
±1kV:
Input/output lines
±2kV: Power
supply lines
±1kV:
Input/output lines
Mains power quality should be that of
a typical commercial or hospital
environment.
Surge
IEC 61000-4-5
±1 kV differential
mode
±2kV common
mode
±1 kV differential
mode
±2kVcommon
mode
Mains power quality should be that of
a typical commercial or hospital
environment.
Voltage drops,
dips, and
fluctuations of
input power
supply line IEC
61000-4-11
<5 %
U
T
(>95 % dip in
U
T)
for 0,5 cycle
40 %
U
T
(60 % dip in
U
T)
for 5 cycles
70 %
U
T
<5 %
U
T
(>95 % dip in
U
T)
for 0,5 cycle
40 %
U
T
(60 % dip in
U
T)
for 5 cycles
70 %
U
T
Mains power quality should be that of
a typical commercial or hospital
environment. If the user of the
ACCUNIQ BP800 requires continued
operation
during
power
mains
interruptions, it is recommended that
the ACCUNIQ BP800 be powered