28
1. Measurement error and solution
ERROR
CAUSE
SOLUTION
Err 0
• When inflation lever is not
sufficient.
• When air hose is not connected
well.
• When air leaks from the cuff.
• When battery power is low.
• Please set higher inflation level.
• Check if there is any ‘CAUSE’.
• After settle it, monitor again.
Err 1
• When a cuff is not connected to
a monitor.
• When a cuff is positioned
incorrectly
• Please check if the cuff is connected well to
the unit and monitor again.
Err 2
• When the pressurization is over
than 300
㎜
Hg.
• Please turn the power off then on and
monitor again.
Err 3
• When monitoring is carried out
the wrong way.
• When a readed data is not
output as a result lf noise or
movement detected.
• Please read ‘wrapping the cuff’ and ‘correct
measuring posture’ in ‘Measurement’ part.
• Monitor again following the way instructed.
Err 4
• When moving or speaking is
sensed during monitoring.
• Please do not move or speak during
monitoring.
Err 5
• When battery is not fixed well.
• When the polarities are
positioned incorrectly.
• Check if the polarities of the batteries are
positioned correctly.
• After disconnecting batteries in unit and press
[I/O] button, then replace them.
TROUBLE SHOOTING