Содержание BP800

Страница 1: ......

Страница 2: ...THE PERSONS RESPONSIBLE FOR PLACING DEVICES ON THE EC MARKET UNDER MDD 93 42 EEC SELVAS Healthcare Inc 29 Gongdan 4 ro Jillyang eup Gyeongsan si Gyeongsangbuk do 38470 Republic of Korea TEL 82 53 856...

Страница 3: ...3 SIDE PART 15 4 ACCESSORIES 16 5 OPTIONS 16 INSTALLATION 17 1 HOW TO PUT BATTERY 17 2 HOW TO CONNECT COMMUNICATION PORT 18 3 SETTING 19 MEASUREMENT 23 1 WRAPPING THE CUFF 23 2 MEASUREMENT PREPARATIO...

Страница 4: ...is designed for use with patients with normal sinus rhythms 4 Measurement accuracy for the ACCUNIQ BP800 ABPM may be affected by position of the subject physical conditions movement and use outside o...

Страница 5: ...5 Safety of Medical Electrical Equipment 2 This device is in compliance with Class A for Noise Emission Level B for Noise immunity by standard of IEC 60601 1 2 2007 Electromagnetic Compatibility Requi...

Страница 6: ...ur company or its authorized dealer for instructions before doing so When you connect computer or other system to the unit RS 232C the attached systems should be those certified by IEC 950 or equivale...

Страница 7: ...S 18 a statement that portable and mobile RF communication equipment can affect MEDICAL ELECTRICAL EQUIPMENT 19 Please consult a physician or a trained health professional for interpretation of measur...

Страница 8: ...Turning on the recorder Turning off the recorder Date of manufacture Manufacturer Non ionizing radiation CE mark Serial No Refer to instruction manual booklet Authorized representative in the European...

Страница 9: ...agnetic immunity The ACCUNIQ BP800 is intended for use in the electromagnetic environment specified below The customer or the user of the ACCUNIQ BP800 should assure that it is used in such an environ...

Страница 10: ...mmunity test IEC 60601 test level Compliance level Electromagnetic environment guidance Conducted RF IEC 61000 4 6 Radiated RF IEC 61000 4 3 3 Vrms 150 kHz to 80 MHz 3 V m 80 MHz to 2 5 GHz 3 Vrms 3 V...

Страница 11: ...ommunications equipment and the ACCUNIQ BP800 The ACCUNIQ BP800 is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled The customer or the user of the A...

Страница 12: ...started with START STOP button Systolic blood pressure This icon will be indicated when wake up time is set Diastolic blood pressure This icon will be indicated when wake up time automatic measuremen...

Страница 13: ...c UP button Number is going up when setting the device DOWN button Number is going down when setting the device Battery cover It is a cover for protecting battery inlet When replace the battery replac...

Страница 14: ...14 3 SIDE PART USB Port This port is for transmitting data to data managing program on PC USB Port...

Страница 15: ...ry AAA X 4 USB cable Carrying bag and strap Data managing program 5 Options Cuff For small adult Cuff For large adult Cuff and connecting hose USB cable Carrying bag and strap Battery AAA X 4 Cuff For...

Страница 16: ...y cover and put each 2 batteries AAA type battery on front and rear part Front Part Rear Part After putting batteries on rear part turn on the ON OFF switch as shown in the picture Note Replace all ba...

Страница 17: ...17 2 How to connect communication port Connect the USB port which located on side part of main unit and computer USB port by USB cable...

Страница 18: ...started setting Month day setting After setting minute press MENU button once again for entering month setting mode Press UP and DOWN button for setting month Press MENU button once again for entering...

Страница 19: ...m After setting the maximal value press the MENU button once again and the device will show SYS and L on the LCD screen You may also set the minimal blood pressure alarm Set the desired value using th...

Страница 20: ...e up time After setting desired wake up time press the MENU button again and the LCD screen will display M and A and enters measuring intervals menu Set desired interval using UP and DOWN button The d...

Страница 21: ...interval using the UP and DOWN button The device will automatically measure the blood pressure at every preset interval and the device operates manually when the value is set at 0 Setting bed time Af...

Страница 22: ...ns which are printed directly on the cuff as it should be applied brachial artery inside forearm Position the lower edge of the cuff is 2 3cm above the elbow joint as shown in the picture Strain the c...

Страница 23: ...ound the arm two finger loose and make sure the cuff is not folded or misplaced Hang the air hose around the neck and put the belt on the flank so that the device is placed on the other side of the cu...

Страница 24: ...at ically measures blood pressure according to the preset value If not press the START button on the device to begin measuring Note If the cuff pressure is too high and it causes pain press the STOP b...

Страница 25: ...he screen Please take a rest and redo the measurement with accurate measuring techniques If the same message is repeatedly displayed even with correct measuring conditions request for a repair service...

Страница 26: ...nit is exposed to high concentration of oil vapor Where the unit is exposed to salty atmosphere Where the unit is exposed to explosive gas or dust Where the unit is exposed to excessive shocks or vibr...

Страница 27: ...Confirm parts of contact with patient directly Keep clean for these parts Instruct the patient on how to stop operation if there is an abnormal measurement and how to remove the cuff if there is exce...

Страница 28: ...on is over than 300 Hg Please turn the power off then on and monitor again Err 3 When monitoring is carried out the wrong way When a readed data is not output as a result lf noise or movement detected...

Страница 29: ...lower than your heart during monitoring Make sure that your arm position at the same level as your heart then monitor it again Re pressurization In case of insufficient inflation The movement of the...

Страница 30: ...e contact our company or distributor periodically for calibration How to contact our company Write us at SELVAS Healthcare Inc 29 Gongdan 4 ro Jillyang eup Gyeongsan si Gyeongsangbuk do 38470 Republic...

Страница 31: ...prox 1 minute Power DC 6V Battery type AAA 1 5V 4 Provided Memory 576 memory Ambience for operation Temperature 10 40 Homidity 30 75 Ambience for storage Temperature 20 60 Humidity Less than 95 Data t...

Страница 32: ...he date of purchase is 1 the date specified in our records if you purchased the product directly from us 2 the date specified in the warranty registration card that we ask you to send to us or 3 if yo...

Страница 33: ...Unit Recorder I D Date Out Date Due In MEASUREMENT PERIOD From To INTERVAL No 1 to every min No 4 to every min No 2 to every min No 5 to every min No 3 to every min No 6 to every min ACTIVITY RECORD T...

Страница 34: ...t indicated on LCD anymore Press START STOP button again and the Automatic measurement starts It is indicated when the asleep time is set It is indicated when the measurement period is set for sleepin...

Страница 35: ...oth operation Pass Fail 2 Recorder Smooth operation with no abnormal sound Pass Fail Mainframe 3 Cuffs Smooth operation Pass Fail Accessories 1 Battery Smooth operation and removal Pass Fail Electrica...

Страница 36: ...peration Pass Fail Mainframe 2 Recorder Smooth operation with no abnormal sound Pass Fail Accessories 1 Battery Smooth operation and removal Pass Fail Electrical Check 1 Battery Screen display upon po...

Страница 37: ...BP800E 20161124 Rev 1 1 EUROPEAN REPRESENTATIVE VITAKO Sp z o o ul Stanis awa aryna 7c 02 593 Warszawa POLAND TEL 48 22 400 8000...