background image

41

40

pH electrode activation (for glass bodied electrodes only)

WARNING:

Only qualified persons proficient with the safe handling of dangerous

chemicals should perform the procedure below. Provide proper containers, fume
hoods, ventilation, and waste disposal. Safety goggles and protective clothing must
be worn while performing this procedure. If possible, replace with another electrode
instead of performing this reactivation procedure.

1. 

Dip or stir the pH electrode

in alcohol for 5 minutes.

2. 

Leave the electrode in tap water

for 15 minutes.

3. 

Dip and stir the electrode

in concentrate acid (such as HCl or H

2

SO

4

for five minutes.

4. 

Repeat step 2.

5. 

Dip and stir in strong base

(NaOH) for five minutes.

6. 

Leave for 15 minutes

in distilled or deionized water.

7. 

Now test with standard calibration buffer solutions

to see if the electrode yields

acceptable results. You may repeat step 3 through 6 up to three times for better
response. If the response does not improve, then your electrode is no longer 
functioning. Replace with a new electrode.

9. Probe Care and Maintenance

Since your pH electrode is susceptible to dirt and contamination, clean it every one
to three months depending on the extent and condition of use.

NOTE:

for specialty electrode care, consult the instruction manual included with
your electrode.

pH electrode storage

For best results, always keep the pH bulb wet. Use the protective electrode storage
bottle or rubber cap filled with electrode storage solution to store your electrode.
Also, you can store in a pH 4 buffer with 1/100 part of saturated KCl. Other pH
buffers are OK for storage, but NEVER use distilled water for storage.

After measuring

1. 

Rinse the pH electrode and reference junction in de-ionized water.

2. 

Store the electrode as recommended above in “pH electrode storage,” 
or as recommended by the manufacturer.

3. 

Prior to next use, rinse the liquid junction with de-ionized water and tap dry—
never wipe electrode.

NOTE:

If this does not restore electrode to normal response, see 
“Reactivating the pH electrode” section below.

pH electrode cleaning

Salt deposits:

dissolve the deposits by immersing the electrode in tap water for ten

to fifteen minutes. Ten thoroughly rinse with distilled water.

Oil/grease film:

wash electrode pH bulb gently in some detergent and water. Rinse

electrode tip with distilled water or use a general purpose electrode cleaner.

Clogged reference junction:

heat a diluted KCl solution to 60 to 80°C. Place the 

sensing part of the electrode into the heated solution for about 10 minutes. Allow 
the electrode to cool in some unheated KCl solution.

Protein deposits:

prepare a 1% pepsin solution in 0.1 M of HCl. Set the electrode in

the solution for five to ten minutes. Rinse the electrode with distilled water.

Reactivating the pH electrode

If stored and cleaned properly, your pH electrode should be ready for immediate
use. However, a dehydrated bulb may cause sluggish response. To rehydrate the
bulb, immerse the electrode in a pH 4 buffer solution for 10 to 30 minutes. If this
fails, the electrode requires activation. Never touch or rub glass bulb. Contact
builds up an electrostatic charge.

Содержание AP71

Страница 1: ...OPERATING INSTRUCTIONS accumet 13636AP71 and 13636AP72 AP71 and AP72 Portable Waterproof pH mV C Meter 1 02 IP 67...

Страница 2: ...s 23 24 7 1 Memory Input 23 7 2 Memory Recall 24 8 Advanced setup functions 25 39 8 1 Advanced setup mode overview 26 27 AP71 meter 26 AP72 meter 27 8 2 Program 1 0 Memory clear 28 8 3 Program 2 0 Vie...

Страница 3: ...ured pH mV or Relative mV reading The secondary display shows the temperature The display also shows error messages keypad functions and program functions 2 1 Primary display 5 6 Secondary display 3 2...

Страница 4: ...e from pH calibration mode see page 17 for directions In advanced set up mode Press CAL MEAS to return to main menu from sub menus Press CAL MEAS again to return to measurement mode from main menu ENT...

Страница 5: ...ure the slots of the connector are in line with the posts of the socket Rotate and push the connector clockwise until it locks See figure below 2 To remove probe push and rotate the connector counterc...

Страница 6: ...first holder See figure 4 Slide the flange of the second holder into the slot of the first holder until the tops of the hold ers are aligned and secure Inserting the Electrode into the Electrode Hold...

Страница 7: ...switch on the meter the meter starts up in the units last used For example if you shut the meter off in mV units the meter will read mV units when you switch the meter on Be sure to remove the protec...

Страница 8: ...g one point calibration go to step 9 7 Rinse the probe with de ionized water or a rinse solution and place it in the next pH buffer 8 Follow steps 5 to 8 for additional calibration points 9 When calib...

Страница 9: ...long period of time or if you have a replacement probe 4 5 B B A A Relative mV Calibration 1 While in the measurement function press MODE to enter the mV mode The mV indicator appears in the upper rig...

Страница 10: ...et temperature Notes To exit this program without confirming the manual temperature compensation value DO NOT press ENTER in step 6 Press CAL MEAS instead Manual Temperature Compensation A A 5 2 5 Mea...

Страница 11: ...tween pH and mV or Rel mV readings press the MODE key 21 20 Taking Measurements Be sure to remove the electrode soaker bottle or protective rubber cap on the electrode before measurement A A 5 3 Takin...

Страница 12: ...han 5 seconds The hold indicator appears Press the HOLD key to release the reading To switch on or off the auto endpoint feature see Set up program P4 1 on page 32 23 22 7 Memory functions Memory Inpu...

Страница 13: ...to the memory location screen AP72 meter only Press the ENTER key again to view the date and time the reading was taken Press the ENTER key again to return to the memory location screen See figure 4...

Страница 14: ...ectrode data In pH measurement mode P3 1 pH electrode offset P3 2 pH electrode slope In mV or relative mV measurement mode P3 1 relative mV offset P1 0 Memory clear P1 0 Clear all stored readings P2 0...

Страница 15: ...mory Clear Use this parameter to clear all memory values when you need to store a new series of values This lets you avoid confusing the old values with the new ones NO is the default setting NOTE Sel...

Страница 16: ...eed to step 4 of P3 2 Or press the CAL MEAS key to return to measurement mode 31 30 P3 0 Viewing electrode data Program 3 has two view only options that lets you check the electrode parameters for dia...

Страница 17: ...Notes Meter default is set for Ready indicator on and auto endpoint function off 33 32 P4 0 Unit configuration P4 1 READY indicator and auto endpoint function Program P4 1 lets you select READY indic...

Страница 18: ...12 45 NIST buffers pH 1 68 4 01 6 86 9 18 12 45 DIN buffers pH 1 09 3 06 4 65 6 79 9 23 12 75 5 Press the ENTER key to confirm selection and to return to the subgroup menu Press the CAL MEAS key to re...

Страница 19: ...ly The AP72 meter features a real time calendar and clock This helps you meet GLP Good Laboratory Practice standards 8 7 A B From measurement mode 1 Press Setup key to enter Set Up mode 2 Press the an...

Страница 20: ...eturn to measurement mode A B 11 Press the ENTER key to confirm the date and move to hour selection The hour digits will flash See figure 12 Press the and keys to toggle to the correct hour Note the A...

Страница 21: ...olution to store your electrode Also you can store in a pH 4 buffer with 1 100 part of saturated KCl Other pH buffers are OK for storage but NEVER use distilled water for storage After measuring 1 Rin...

Страница 22: ...y eliminate the error message See figure If error persists or the meter shows incorrect values return the meter For a complete diagram of the display see page 5 ERR 1 in primary display A A 10 Trouble...

Страница 23: ...USA buffers pH 1 68 4 01 7 00 10 00 12 45 NIST buffers pH 1 68 4 01 6 86 9 18 12 45 DIN buffers pH 1 09 3 06 4 65 6 79 9 23 12 75 Temperature Offset in 0 1 C increments mV Offset up to 150 mV Memory A...

Страница 24: ...47 46...

Отзывы: