background image

1. 

If necessary, press the MODE key to
select pH mode

. The pH indicator 

appears in the upper right hand corner 
of the display.

See figure 

2. 

Rinse the probe thoroughly with 
de-ionized water or a rinse solution.

Do

not wipe the probe; this causes a build-up
of electrostatic charge on the glass surface.

3. 

Dip the probe into the calibration buffer.
The end of the probe must be completely
immersed into the sample. Stir the probe
gently to create a homogeneous sample.

4. 

Press CAL/MEAS

to enter pH calibration

mode. The CAL indicator lights. The 
primary display will show the measured
reading while the smaller secondary dis-
play will indicate the pH standard buffer
solution. 

See figure 

5. 

Wait for the measured pH value to 
stabilize.

See figure 

15

14

6. 

Press ENTER to confirm calibration

The meter is now calibrated to the 
current buffer.

See figure 

• If you are performing multipoint 

calibration, go to step 7. 

• If you are performing one-point 

calibration, go to step 9.

7. 

Rinse the probe with de-ionized water
or a rinse solution, and place it in the
next pH buffer. 

8. 

Follow steps 5 to 8 for additional 
calibration points. 

9. 

When calibration is complete, 
press CAL/MEAS to return to pH 
measurement mode

.

Notes

To exit from pH Calibration mode without
confirming calibration,

DO NOT press

ENTER 

in step 6. Press CAL/MEAS instead.

If the selected buffer value is not within
±1.0 pH from the measured pH value: the
electrode and buffer icon blink and the ERR
annunciator appears in the lower left corner
of the display.

To limit the number of pH buffer values
available during calibration, see section P4.2
on page 33.

D

C

D

B

C

B

A

A

pH calibration

NOTE: 

We recommend that you perform at least a 2-point calibration using standard
buffers that bracket (one above and one below) the expected sample range. 
You can perform a 1-point calibration, but make sure that the buffer value is
close to the sample value you are measuring.

4.3

C

If the READY indicator has been activiated
(set up program P4.1—see page 32), the READY
annunciator lights when the reading is stable.

Содержание AP71

Страница 1: ...OPERATING INSTRUCTIONS accumet 13636AP71 and 13636AP72 AP71 and AP72 Portable Waterproof pH mV C Meter 1 02 IP 67...

Страница 2: ...s 23 24 7 1 Memory Input 23 7 2 Memory Recall 24 8 Advanced setup functions 25 39 8 1 Advanced setup mode overview 26 27 AP71 meter 26 AP72 meter 27 8 2 Program 1 0 Memory clear 28 8 3 Program 2 0 Vie...

Страница 3: ...ured pH mV or Relative mV reading The secondary display shows the temperature The display also shows error messages keypad functions and program functions 2 1 Primary display 5 6 Secondary display 3 2...

Страница 4: ...e from pH calibration mode see page 17 for directions In advanced set up mode Press CAL MEAS to return to main menu from sub menus Press CAL MEAS again to return to measurement mode from main menu ENT...

Страница 5: ...ure the slots of the connector are in line with the posts of the socket Rotate and push the connector clockwise until it locks See figure below 2 To remove probe push and rotate the connector counterc...

Страница 6: ...first holder See figure 4 Slide the flange of the second holder into the slot of the first holder until the tops of the hold ers are aligned and secure Inserting the Electrode into the Electrode Hold...

Страница 7: ...switch on the meter the meter starts up in the units last used For example if you shut the meter off in mV units the meter will read mV units when you switch the meter on Be sure to remove the protec...

Страница 8: ...g one point calibration go to step 9 7 Rinse the probe with de ionized water or a rinse solution and place it in the next pH buffer 8 Follow steps 5 to 8 for additional calibration points 9 When calib...

Страница 9: ...long period of time or if you have a replacement probe 4 5 B B A A Relative mV Calibration 1 While in the measurement function press MODE to enter the mV mode The mV indicator appears in the upper rig...

Страница 10: ...et temperature Notes To exit this program without confirming the manual temperature compensation value DO NOT press ENTER in step 6 Press CAL MEAS instead Manual Temperature Compensation A A 5 2 5 Mea...

Страница 11: ...tween pH and mV or Rel mV readings press the MODE key 21 20 Taking Measurements Be sure to remove the electrode soaker bottle or protective rubber cap on the electrode before measurement A A 5 3 Takin...

Страница 12: ...han 5 seconds The hold indicator appears Press the HOLD key to release the reading To switch on or off the auto endpoint feature see Set up program P4 1 on page 32 23 22 7 Memory functions Memory Inpu...

Страница 13: ...to the memory location screen AP72 meter only Press the ENTER key again to view the date and time the reading was taken Press the ENTER key again to return to the memory location screen See figure 4...

Страница 14: ...ectrode data In pH measurement mode P3 1 pH electrode offset P3 2 pH electrode slope In mV or relative mV measurement mode P3 1 relative mV offset P1 0 Memory clear P1 0 Clear all stored readings P2 0...

Страница 15: ...mory Clear Use this parameter to clear all memory values when you need to store a new series of values This lets you avoid confusing the old values with the new ones NO is the default setting NOTE Sel...

Страница 16: ...eed to step 4 of P3 2 Or press the CAL MEAS key to return to measurement mode 31 30 P3 0 Viewing electrode data Program 3 has two view only options that lets you check the electrode parameters for dia...

Страница 17: ...Notes Meter default is set for Ready indicator on and auto endpoint function off 33 32 P4 0 Unit configuration P4 1 READY indicator and auto endpoint function Program P4 1 lets you select READY indic...

Страница 18: ...12 45 NIST buffers pH 1 68 4 01 6 86 9 18 12 45 DIN buffers pH 1 09 3 06 4 65 6 79 9 23 12 75 5 Press the ENTER key to confirm selection and to return to the subgroup menu Press the CAL MEAS key to re...

Страница 19: ...ly The AP72 meter features a real time calendar and clock This helps you meet GLP Good Laboratory Practice standards 8 7 A B From measurement mode 1 Press Setup key to enter Set Up mode 2 Press the an...

Страница 20: ...eturn to measurement mode A B 11 Press the ENTER key to confirm the date and move to hour selection The hour digits will flash See figure 12 Press the and keys to toggle to the correct hour Note the A...

Страница 21: ...olution to store your electrode Also you can store in a pH 4 buffer with 1 100 part of saturated KCl Other pH buffers are OK for storage but NEVER use distilled water for storage After measuring 1 Rin...

Страница 22: ...y eliminate the error message See figure If error persists or the meter shows incorrect values return the meter For a complete diagram of the display see page 5 ERR 1 in primary display A A 10 Trouble...

Страница 23: ...USA buffers pH 1 68 4 01 7 00 10 00 12 45 NIST buffers pH 1 68 4 01 6 86 9 18 12 45 DIN buffers pH 1 09 3 06 4 65 6 79 9 23 12 75 Temperature Offset in 0 1 C increments mV Offset up to 150 mV Memory A...

Страница 24: ...47 46...

Отзывы: