AccuMed CMS-50D Скачать руководство пользователя страница 13

Pulse Parameter Specification 

Measuring range 

30bpm

240bpm, (the resolution is 1bpm) 

Accuracy 

±2bpm or

±

2% (select larger) 

 

Pulse Intensity 

Range 

Continuous bar-graph display, the higher display indicate the stronger 

pulse. 

Battery Requirement 

1.5V (AAA size) alkaline batteries × 2   

Battery useful life

 

The two batteries can work continually 30 hours. 

 

Dimensions and Weight 

Dimensions 

57(L) × 31(W) × 32(H) mm 

Weight 

About 50g (with the batteries) 

12. Key of Symbols 

 

Symbol 

Description 

 Type 

BF 

 

Warning – See User Manual 

 

The pulse oxygen saturation(%) 

 

Pulse rate (bpm) 

 

Low-voltage  

SN Serial 

number 

 

Battery anode 

 

Battery cathode 

 

10

Содержание CMS-50D

Страница 1: ...l includes the Fingertip Pulse Oximeter s features and requirements main structure functions specifications correct methods for transportation installation usage operation repair maintenance and stora...

Страница 2: ...ce does not cover such faults This product is medical device and can be used repeatedly Its using life is 3 years WARNING 0 Uncomfortable or painful feeling may appear if using the device ceaselessly...

Страница 3: ...CAUTION 3 3 1 PRINCIPLE OF MEASUREMENT 3 3 2 THE ATTENTION OF OPERATION 3 3 3 CLINICAL RESTRICTIONS 4 4 TECHNICAL SPECIFICATIONS 4 5 ACCESSORIES 5 6 INSTALLATION 5 6 1 VIEW OF THE FRONT PANEL 5 6 2 M...

Страница 4: ...avoid device damage Please don t use hanging rope if allergic to hanging rope 0 Please don t use this product if allergic to rubber cushion 0 Please dispose the device accessory and packing including...

Страница 5: ...in the range from 70 to 100 2 1 Features A The device can accurately measure SpO2 B The device will power off automatically if no signal is detected within 5 seconds C Two AAA size batteries are requi...

Страница 6: ...unctional O2Hb saturation O2Hb RHb O2Hb x 100 Present SpO2 oximeter transmits light of two wavelengths only red light wavelength 660 nm and infrared wavelength 940nm to differentiate HbO2 from HbR One...

Страница 7: ...t body temperature major bleeding or use of vascular contracting drug the SpO2 waveform PLETH will decrease In this case the measurement will be more sensitive to interference B For those with a subst...

Страница 8: ...se saturation is at 0 4 Accuracy of SpO2 4 Accuracy of Pulse rate 2bpm or 2 select larger G Resistance to surrounding light The difference between the value measured in the condition of man made light...

Страница 9: ...gure 3 And insert the two AAA size batteries properly in the right direction according to the and symbol 2 Replace the cover Please be cautious when you insert the batteries for improper insertion may...

Страница 10: ...clip as shown in Figure 5 Figure5 Put finger in position C Let the patient s finger put into the rubber cushions of the clip make sure the finger is in the right position and then clip the finger a Pr...

Страница 11: ...ong time 4 The best storage environment of the device is 20 C to 55 C ambient temperature and not higher than 95 relative humidity Please maintain properly for ensuring the device can be used normally...

Страница 12: ...e center The display is off suddenly 1 The device will power off automatically if no signal is detected within 5 seconds 2 The batteries are almost drained 1 Normal 2 Change batteries 11 Function Spec...

Страница 13: ...stronger pulse Battery Requirement 1 5V AAA size alkaline batteries 2 Battery useful life The two batteries can work continually 30 hours Dimensions and Weight Dimensions 57 L 31 W 32 H mm Weight Abou...

Страница 14: ...Alarm inhibit 11...

Отзывы: