![Acculab ATL-224 Скачать руководство пользователя страница 54](http://html1.mh-extra.com/html/acculab/atl-224/atl-224_operating-instructions-manual_3812358054.webp)
54
Verified Models with EC-type Approval: Specifications
Model
ATL-6202-V
ATL-4202-V
ATL-2202-V
Type
BD ED 200
BD ED 200
BD ED 200
Accuracy class
1
)
K
K
K
Weighing capacity, Max
1
)
g
6200
4200
2200
Scale interval d
1
)
g
0.01
0.01
0.01
Tare range (subtractive)
<
100% of the maximum capacity
Verification scale interval e
1
)
g
0.1
0.1
0.1
Minimum capacity, Min
1
)
g
0.5
0.5
0.5
Response time (average)
s
1.1
1.1
1.1
Range of use according to CD
1
)
g
0.5 – 6200
0.5 – 4200
0.5 – 2200
Allowable ambient
operating temperature
°C
+10 to +30 (+50°F to +86°F)
Net weight, approx.
kg
3.5
Weighing pan size
mm
180
+
180
Dimensions (W
+
D
+
H)
mm
230
+
303
+
91
Model
ATL-822-V
ATL-8201-V
ATL-6201-V
Type
BD ED 200
BD ED 200
BD ED 200
Accuracy class
1
)
K
K
K
Weighing capacity, Max
1
)
g
820
8200
6200
Scale interval d
1
)
g
0.01
0.1
0.1
Tare range (subtractive)
<
100% of the maximum capacity
Verification scale interval e
1
)
g
0.1
1
1
Minimum capacity, Min
1
)
g
0.5
5
5
Response time (average)
s
1.1
1
1
Range of use according to CD
1
)
g
0.5 – 820
5 – 8200
5 – 6200
Allowable ambient
operating temperature
°C
+10 to +30 (+50°F to +86°F)
Net weight, approx.
kg
3.5
Weighing pan size
mm
180
+
180
Dimensions (W
+
D
+
H)
mm
230
+
303
+
91
1
) CD = Council Directive 90/384/EEC for non-automatic weighing instruments; applicable to the European Economic Area