33
Display symbol:
V
Purpose
Use this program to determine the weights of unstable
samples (e.g., live animals) or to determine weights under
unstable ambient conditions. With this program, the bal-
ance calculates the weight as the average of a defined
number of individual weighing operations (also referred to
as “subweighing operations”).
Preparation
§
Select the Animal weighing application in the menu: see
“Configuration.”
§
Set the following parameters:
Applic.
Application program
AnimalW.
Animal weighing
activty.
Animal activity
Calm
Stable readout
o
Active
Unstable
V.Active
Very unstable readout
Start
manual
o
Autom.
Automatic
o = Factory setting
Changing the Number of Subweighing
Operations
Activate function:
Press the (MENU)* key
Select the desired number of
measurement (1 to 100):
In increments of 1:
Press the(MENU)* key briefly
In increments of 10:
Press and hold the (MENU)* key.
The selected number of measurements is stored in
battery-backed memory
Printout: Animal weighing
mDef
20
: Number of subweighing
operations
x-Net + 410.1 g
: Calculated average
* On some models, the keypad overlay shows the German
word “MENUE“ for MENU.
Animal Weighing/Averaging