Error codes are shown on the main display for approx. 2 seconds.
The program then returns automatically to the previous mode.
Display
Cause
Solution
No segments appear
No AC power is available
Check the AC power supply
on the display
The power supply is not
Plug in the power supply
plugged in
high
The load exceeds the balance capacity
Unload the balance
low
or
Err 54
Something is touching
Move the object that is
the weighing pan
touching the weighing pan
err 54
, typical
Weighing system defect
Contact Acculab dealer
App.err.
Cannot store data:
Increase load
Load on weighing pan too light or no
sample on pan while application is active
dis.err.
Data output not compatible
Change the configuration in
with output format
the operating menu
prt.err.
Interface port for printer
Reset the menu factory settings,
output is blocked
or
Contact Acculab dealer
err 02
Calibration parameter
Calibrate only when zero
not met; e.g.:
is displayed
– balance not tared
– Press [Tare] to tare the balance
– load on weighing pan
– Unload the balance
err 10
The [Tare] key is blocked when
Press [Clear] to clear the tare
there is data in the second tare
memory and release the tare key
memory (net-total); only 1 tare
function can be used at a time
err 11
Tare memory not allowed
Press [Tare]
The weight readout
Unstable ambient conditions
Set up the balance in another area
changes constantly
(excessive vibration or draft)
at the place of installation
A foreign object is caught between
Remove the foreign object
weighing pan and balance housing
The weight readout
The balance was not
Calibrate/adjust the balance
is obviously wrong
calibrated/adjusted
Balance not tared
Tare or zero the balance
before weighing
before weighing