background image

 

5.  Operate system until normal solution tank level is achieved.  Switch system 

off. 

6.  Close inlet and outlet isolation valves. 
7.  Very carefully, pour the acid into the upper chlorinator to dissolve scale. It may 

be necessary to pour directly on the inside walls to remove any scale above the 
water level.  

Always add acid to water, never add water to acid.

 

8.  Let soak for 30 minutes.  As acid dissolves scale, carbon dioxide will be re-

leased, and foaming will occur. If any tablets were left in the feeder, 

chlorine 

gas may also be released.

 

9.  Turn system on to start acid circulation through the blue tubing.  Circulate for 

10 minutes.  

10. Open all valves to begin purging acid from system.  
11. Wash the walls and sieve plate of the feeder with a hose to remove all acid 

residue.  Chlorine gas may be released if any acid remains in the system. 

12. Allow feeder to flush water for 30 minutes. 
13. Turn system off.  System is ready for operation. 
 

Troubleshooting 

System overflows. 

1.  Debris in the float valve.   
2.  Debris in the check valve.  
3.  Float valve stuck in down position.  Clean float valve. 

Pump will not run. 

1.  Check power supply and connections, and controller 110v signal.   
2.  Check for 110 voltage across terminals #2 and #5.  If voltage present, wiring 

or pump motor failure. 

Pump loses prime (solution tank drains) 

1.  Maximum inlet flow exceeded.  See Start-Up instructions—point #5. 

 

Содержание PowerBase 3070AT

Страница 1: ...ets only Do not mix Cal Hypo tablets with stabilized chlorine or bromine products or other sanitizing chemicals Fire or explosion could result Rev 04 26 10 NSF 50 The PowerBase 3070AT Chlorinator is N...

Страница 2: ...or use ONLY with PPG Accu Tab Blue SI tablets utilizing the PPG patented erosion feeder technology that erodes only the bottom layer of tablets on the sieve plate while keeping tablets above the sieve...

Страница 3: ...e chlorine application point with 1 1 2 PVC pipe or hose Include an isolation valve in the line 9 Plug the power cord into an 110v receptacle 10 If a chemical controller is utilized plug the controlle...

Страница 4: ...4...

Страница 5: ...maximum inlet flow is determined 6 Adjust flow through feeder using the discharge gate valve The chlorine delivery is determined by the flow rate through the chlorinator Use the graph above to determ...

Страница 6: ...on system may develop a build up of scale This should be cleaned on a periodic basis up to twice per year If scale formation is severe more frequent cleanings may be necessary Cleaning may be accompli...

Страница 7: ...hrough the blue tubing Circulate for 10 minutes 10 Open all valves to begin purging acid from system 11 Wash the walls and sieve plate of the feeder with a hose to remove all acid residue Chlorine gas...

Страница 8: ...8 Electrical Drawing Base System...

Страница 9: ...mpers between Terminal 1 4 Terminal 2 6 Electrical Drawing With HHL Pump Switch If a chemical controller is NOT used with the HHL Pump Switch the jumpers should be placed between Terminals 1 and 4 Ter...

Страница 10: ...talled and operated and maintained in accordance with this manual and the instructions ac companying the System iii you at your cost promptly return the System or defective part to such location as ma...

Страница 11: ...11...

Отзывы: