EXC-350 MICROSCOPE SERIES
ACCU-SCOPE
®
15
ADJUSTMENT
& OPERATION
(continued)
Replacing the Fuses
Turn the main switch to “0” (OFF) before
replacing a fuse. Unplug the power cord from the
microscope.
Screw off the fuse
①
or
②
from the fuse base with a
Phillip’s head type screwdriver.
Install a new fuse and screw it on the fuse base
(Fig. 19).
Specification of the fuse: 250V, 3.15A.
CAT #350-3277-3
Installation of a Camera Adapter
Loosen the lock screw
①
on the trinocular head and
remove the dust cap
②
on the trinocular port.
Remove the dust caps of the C-mount adapter
and
insert the C-mount into the trinocular head as shown in
Fig. 20.
Tighten the lock screw
①
.
Loosen the lock screw
④
on the C-mount.
Remove the C-mount
⑤
from the lower portion of the
adapter.
Screw onto the camera and then install assembled
C-mount and camera back onto the lower assembly.
Lock it in place by tightening the lock screw.
Observe the image through the eyepieces and bring it
into focus.
Look at the camera image and adjust the focus of the
camera image via the rotating ring
③
until it matches
the focus as seen through the eyepieces.
Fig. 19
①
②
Fig. 20
①
②
③
④
⑤