EXC-350 MICROSCOPE SERIES
ACCU-SCOPE
®
12
ADJUSTMENT
& OPERATION
(continued)
Adjusting the Focusing Tension
If the handle is very heavy when focusing or the specimen
leaves the focus plane after focusing or the stage lowers by
itself, please adjust the tension adjustment ring
①
.
Located
on the left side of the stand between the coarse adjustment
knob and the vertical arm is an adjustable tension control
dial that is preset at our facility. This allows the user to
adjust the coarse control tension to their individual
preference. (Fig. 9).
Adjusting the Diopter
Set the diopter rings
①
&
②
o
f both eyepieces to “0” (zero)
position. (Do this when users change, because different
users will have different diopter settings.)
Place an easy-to-observe specimen on the stage plate, i.e.,
a coin. Looking only into the left eyepiece, adjust the diopter
ring on left eyepiece to focus the specimen. Then repeat
procedure for the right eyepiece. Different users will have
different diopter settings, so adjust the diopter for each
individual user.
Be sure to make a note of your eye’s diopter
setting should another user change the setting. (Fig. 10)
Adjusting the Interpupillary Distance
To adjust the interpupillary distance, hold the left and right
eyetubes while observing a specimen. Rotate the eyetubes
around the central axis until the fields of view of both
eyetubes coincide completely. A complete circle should be
seen in the viewing field when viewing the specimen slide.
An improper adjustment will cause operator fatigue and will
disrupt the objective parfocality.
Where
“
●
”
①
on the eyepiece tube lines up to the
interpupillary scale
②
, then that is the number for the
interpupillary distance. Range
:
50
~
75mm.
(Fig. 11).
Remember your interpupillary for future operation.
NOTE:
The eyepiece tubes can be rotated 180° to increase
the eyepoint height by 34mm to accommodate the needs of
different users. (Fig. 12)
Fig. 9
Fig. 10
Fig. 11
Fig. 12
②
①
①
①
②