239
Technical data 18
Temperature
and humidity
during testing
during storage
+10 to +32 °C at 15 to 85 % relative humidity, including the following extreme
conditions:
+32 °C at 85 % relative humidity and +40 °C at 45 % relative humidity.
15 to 93 % relative humidity
Meter without batteries and without test cassette: -25 to +70 °C
” with batteries and without test cassette:
-10 to +50 °C
” with batteries and with test cassette:
+2 to +30 °C
Altitude
Sea level to 4000 m
Memory
500 results with time and date, averages for 7, 14 and 30 days
Dimensions
123 × 66 × 28 mm with lancing device
Weight
approx. 150 g with lancing device, batteries, test cassette and lancet drum
Display
oLed (organic Light emitting diode) display
Interface
Infrared window
Safety class
III
LED/IRED
Class 1