Acoustic mode 6
6.4 Announcement of saved test results
only the last (most recent) of the saved test results is
announced acoustically.
The result is announced three times in succession. The
announcement follows the same pattern as the an-
nouncement of a result immediately after the test. you
will hear: result result result.
If you directly retrieve saved results using the button
, you will at first hear the power-on beep and im
-
mediately afterwards the result:
result result result.
Flags are not acoustically announced. If you flag the
last saved result with
(General)
,
(Before food)
or
(After food)
, this result will be announced like a
non-flagged result.
If xx.x or x.x is displayed as the last (most recent)
result, the last test did not deliver any result (see
Chapter 5.1). In this case, the sounded beeps are
identical to those for an error message (see next page).
There is no acoustic announcement when
the last saved result is a control test flagged with
,
no results are saved,
you retrieve an average.