Chapter 1: Your New Meter
ACCU-CHEK Customer Service Center 1-800-858-8072
Understanding Control Test Results
Chapter 2: Control Testing
ACCU-CHEK Customer Care Service Center 1-800-858-8072
Example
The label on your test drum vial shows the acceptable ranges for
low and high test solutions. Your control test results should be
within the listed range. Make sure you compare the result to the
range for the level of control solution you used (Low or High).
When a control test result is within the acceptable range, your test
drum and meter are working properly.
Range (mg/dL)
Low
33–63
High
311–421
Low
43–73
High
321–431