Performing a Control Test
You need the meter, a strip, and control solution Level 1 and/or Level 2.
The control level is printed on the bottle label.
1. Insert the strip into the meter in
the direction of the arrows. The meter
turns on.
2. Make sure the code number on the
display matches the code number on
the strip container. If you miss
seeing the code number, remove the
strip and reinsert it into the meter.
53
Содержание Aviva Nano
Страница 1: ...BLOOD GLUCOSE METER GLYCOM TRE User s Manual Manuel de l utilisateur ACCU CHEK Aviva Nano...
Страница 5: ...4...
Страница 13: ...12...
Страница 51: ...50 50...
Страница 73: ...72...