120
Technical data
Performance analysis
The performance data for the Accu-Chek Active system (Accu-Chek Active
meter with Accu-Chek Active test strips) was obtained using capillary blood
from diabetic patients (method comparison, accuracy), venous blood (repeat-
ability) and control solution (reproducibility). The system is calibrated with
venous blood containing various glucose concentrations. The reference values
for this were obtained using the hexokinase method. For method comparison,
the results were compared with results obtained using the hexokinase method
with deproteination (automatic analyser). The hexokinase method is traceable to
a NIST standard.
The Accu-Chek Active system meets the ISO 15197 requirements.