81
Checking your meter
Did you compare the test result with the
concentration data that corresponds to the
control solution you used?
yes
Is the concentration table on the test strip
container from which you removed the test
strip?
yes
Is the test strip guide and measuring window
clean?
yes
Has the control solution been open for less
than 3 months?
Once opened, control solutions are stable for
only 3 months and must not be used after
this period.
yes
Did you observe the proper storage condi-
tions for the meter, the test strips and the
control solution (see Chapter 11 and the
package inserts)?
yes
Has the expiry date of the test strips or the
control solution elapsed see the label on the
test strip container (next to the
symbol
and the label on the bottle next to the
symbol and eventually the
symbol on
the display of the meter)?
no
If your meter has been dropped, it can also lead to
implausible results or error messages. In this case
also contact your customer support and service
centre.
Содержание ACTIVE
Страница 123: ...123 Patents Patents 16 US 5 366 609 US 5 424 035 US 5 463 467 US 5 889 585 US 6 055 060 US 6 906 802 ...