SDI-12 Command Summary: 

Acclima SDI-12 sensors implement commands that comply with versions 1.0, 1.1, 1.2, 1.3 and 1.4 
of the SDI-12 specifications.  All commands required for full compliance of the version 1.4 

specifications are implemented in the TDR sensors.  However, the ‘additional measurements’ 

commands in the SDI-12 specification are meaningless to the Acclima SDI-12 sensors, since the 
permittivity, moisture, conductivity, and temperature measurements are all required and all made 
with every single measurement.  Hence the response for all of these additional measurements 
commands is 

“a<CR><LF>”

 as required by the SDI-12 specification.  The sensor ignores 

unimplemented commands outside those required by the SDI-12 specification.  There is no response 
to them. 

Additional ‘Extended Commands’ that are not enumerated here, are in

corporated by Acclima for 

digitized waveform dumps and factory calibration settings and readings. 

 

 

 

 

Command Reference: 

The table below documents all commands supported by the SDI-12 sensor in alphabetical order: 

Command 

Function 

Sensor Response 

?! 

Address Query 

a Note: only one device can be connected to the SDI-12 port 
when this command is used.   a<CR><LF> 

a! 

Acknowledge 
Active 

a      a<CR><LF> 

aAb! 

Change Address 

b     b<CR><LF>

 

 

aC! 

 

 

aCC! 

Start Concurrent 
Measurement 

 

Start Concurrent 
Measurement-
Request CRC 

 

a00105<CR><LF> 

Measurement takes 1 seconds. 5 values are returned

 

 

aC1! 

aC9! 

Start Additional 
Concurrent 
Measurement 

a00000<CR><LF> 

No data to be provided

 

Содержание TDR Series

Страница 1: ...Acclima TDR Sensor User Manual For Sensor Models TDR305H TDR310H TDR315H...

Страница 2: ...the medium though which the wave has passed The fact that propagation time is independent of soil electrical conductivity leads to the fact that the calculated permittivity is also independent of soi...

Страница 3: ...compacted to avoid air pockets near the waveguide and to prevent the creation of a preferential path for surface water to percolate down to the sensor When the installation is complete the entire soil...

Страница 4: ...was developed from the TDR310H as a measurement solution for soils with exceedingly high salt levels It provides credible measurements in soils with saturated paste extraction EC levels up to 20 dS m...

Страница 5: ...nsor is connected to This may require the use of a laptop and special software provided by the data recorder manufacturer Connecting the Sensors to the Data Recorder The wiring diagram for the sensor...

Страница 6: ...If the data recorder does not supply power to the SDI 12 port then use the diagram below for connecting the sensors...

Страница 7: ...ation The sensor ignores unimplemented commands outside those required by the SDI 12 specification There is no response to them Additional Extended Commands that are not enumerated here are incorporat...

Страница 8: ...aborted Data buffer is empty a CR LF aM Fill data buffer with standard measurement a VVV V TT T PP P EEEE CCCC CR LF where VVV V Volumetric water TT T Temperature C PP P Permittivity no units EEEE Bul...

Страница 9: ...Same as above aV Verify Sensor operation This command produces the same result as the aM Command Same data as aM aD1 aD9 No data for these commands al Send Identification aSSVVVVVVPPPPPPcccxx xx a13A...

Страница 10: ...sensor which is then followed by the serial number Making Measurements with the SDI 12 Sensor There are two commands that cause the SDI 12 sensor to take measurements and store them for subsequent re...

Страница 11: ...or and the commands used to retrieve them Data Item Request Command Response Example Units Volumetric Water Content aD0 25 0 Soil Temperature aD0 32 1 C Bulk Relative Permittivity aD0 32 1 Soil Electr...

Страница 12: ...of seconds that are required to take the readings 001 1 second then finished with the number of data items that will be returned with each measurement request 5 Every response from the sensor is termi...

Страница 13: ...uests the five data items exactly in the same manner as with non concurrent readings Data Communication Error Checking So far all Start Measurement commands that we have discussed have requested data...

Страница 14: ...for sensor operation The measure error code indicates that a point was reached in the waveform analysis where an unresolvable mathematical condition was reached This would be due to shorted rods or t...

Страница 15: ...is desired it is necessary to convert the time ordinate to 5ps units or 220 This number is then converted to hexadecimal 220 decimal 0DC hex If the device address is 1 then the following command is s...

Страница 16: ...Reporting Accuracy 0 25 C 0 to 50 C EC Reporting Accuracy 25 S cm 0 to 1 000 S cm BEC 2 5 1 000 S cm to 6 000 S cm BEC Architectural Characteristics Waveguide Length 305H 310H 315H 5 cm 305H 10 cm 310...

Отзывы: