background image

glued into the end of a pipe to act as an installation handle.  The cable feeds up through the pipe and 
emerges through a notch at the upper end. The pipe can then be capped off at the upper end to prevent 
water ingress.  The pipe/sensor assembly can them be inserted into a hole that was drilled using a 34 
mm flat-bottomed auger.  A mallet can be used to pound the top of the pipe and force the sensor 
waveguide into the undisturbed soil at the bottom of the hole. With the flat-bottomed hole the 
sideways bias on the waveguide rods is minimal thus allowing the rods to enter the soil in a parallel 
alignment. This method of installation is generally easier than the trench method used with the 
TDR315H.  

TDR310H Optimum Application: 

All types of soils are suitable environments for TDR310H measurements. Its sampling volume is only 
about ¼ of the TDR315H but that is the tradeoff for easier installation.  In addition the TDR310H has a 
lower impedance waveguide with a shorter length.  This makes it particularly suited to soils with high 
salt content 

 up to 15 dS/m saturated paste extraction.  

 

TDR305H Description, Installation and Optimum Application: 

The TDR305H was developed from the TDR310H as a measurement solution for soils with exceedingly 
high salt levels.  It provides credible measurements in soils with saturated paste extraction EC levels up 
to 20 dS/m.  Because of its low sampling volume the TDR310H or TDR315H would be better choices for 
applications where salt is not a problem.  

 

Sensor Setup and Connection: 

Acclima’s TDR sensors use the industry standard SDI

-12 communications protocol and are compatible 

with any Data Recorder, Reader or Wireless Device that incorporates a compliant SDI-12 Port. The 
sensor cable includes 3 wires. The red wire is the power line. The blue wire is the bi-directional data line 
and the white wire is the common for both the data and power.  

When sensors are installed to a data recorder or other reading device there are two steps that must be 
taken to insure successful communications. The first is the hardware connection. The second is the 
sensor addressing.  The SDI-12 protocol uses address to distinguish between multiple sensors.  These 
must be set uniquely for each sensor wired to a common SDI-12 recorder or reader.  Valid addresses are 
the numerals 0 through 9, the lower case characters a through z and the upper case characters A 
through Z for total of 62 possible addresses.  The sensors are shipped from the factory with the default 

address ‘0’. 

 

Setting Sensor Addresses: 

Содержание TDR Series

Страница 1: ...Acclima TDR Sensor User Manual For Sensor Models TDR305H TDR310H TDR315H...

Страница 2: ...the medium though which the wave has passed The fact that propagation time is independent of soil electrical conductivity leads to the fact that the calculated permittivity is also independent of soi...

Страница 3: ...compacted to avoid air pockets near the waveguide and to prevent the creation of a preferential path for surface water to percolate down to the sensor When the installation is complete the entire soil...

Страница 4: ...was developed from the TDR310H as a measurement solution for soils with exceedingly high salt levels It provides credible measurements in soils with saturated paste extraction EC levels up to 20 dS m...

Страница 5: ...nsor is connected to This may require the use of a laptop and special software provided by the data recorder manufacturer Connecting the Sensors to the Data Recorder The wiring diagram for the sensor...

Страница 6: ...If the data recorder does not supply power to the SDI 12 port then use the diagram below for connecting the sensors...

Страница 7: ...ation The sensor ignores unimplemented commands outside those required by the SDI 12 specification There is no response to them Additional Extended Commands that are not enumerated here are incorporat...

Страница 8: ...aborted Data buffer is empty a CR LF aM Fill data buffer with standard measurement a VVV V TT T PP P EEEE CCCC CR LF where VVV V Volumetric water TT T Temperature C PP P Permittivity no units EEEE Bul...

Страница 9: ...Same as above aV Verify Sensor operation This command produces the same result as the aM Command Same data as aM aD1 aD9 No data for these commands al Send Identification aSSVVVVVVPPPPPPcccxx xx a13A...

Страница 10: ...sensor which is then followed by the serial number Making Measurements with the SDI 12 Sensor There are two commands that cause the SDI 12 sensor to take measurements and store them for subsequent re...

Страница 11: ...or and the commands used to retrieve them Data Item Request Command Response Example Units Volumetric Water Content aD0 25 0 Soil Temperature aD0 32 1 C Bulk Relative Permittivity aD0 32 1 Soil Electr...

Страница 12: ...of seconds that are required to take the readings 001 1 second then finished with the number of data items that will be returned with each measurement request 5 Every response from the sensor is termi...

Страница 13: ...uests the five data items exactly in the same manner as with non concurrent readings Data Communication Error Checking So far all Start Measurement commands that we have discussed have requested data...

Страница 14: ...for sensor operation The measure error code indicates that a point was reached in the waveform analysis where an unresolvable mathematical condition was reached This would be due to shorted rods or t...

Страница 15: ...is desired it is necessary to convert the time ordinate to 5ps units or 220 This number is then converted to hexadecimal 220 decimal 0DC hex If the device address is 1 then the following command is s...

Страница 16: ...Reporting Accuracy 0 25 C 0 to 50 C EC Reporting Accuracy 25 S cm 0 to 1 000 S cm BEC 2 5 1 000 S cm to 6 000 S cm BEC Architectural Characteristics Waveguide Length 305H 310H 315H 5 cm 305H 10 cm 310...

Отзывы: