12                                                                                                                                      OMNISOUND

®

 3000C PRO USER MANUAL 

 

 

 

COPYRIGHT © 2001 - 2017, ACCELERATED CARE PLUS CORP., ALL RIGHTS RESERVED 

Transducer Controls / Operation 
 

The  Omnisound

®

  3000C  Pro  transducers  are  available  in  two  sizes,  2  cm

2

  and  5  cm

2

.    Each  transducer  may  be 

operated  at  1  or  3  MHz  operating  frequencies,  which  control  the  depth  of  penetration.    A  unique  feature  of 
Omnisound

®

  3000C  Pro  transducers  is  the  fully  adjustable  transducer  angle.    To  adjust  transducer  angle  hold  the 

transducer  in  one  hand  while  adjusting  the  transducer  head  to  the  desired  position.    The  transducer  should  not  be 
stressed  when it reaches the end of its  movement range to avoid damage of the rotary joint.   ACP transducers are 
fully  waterproof  and  may  be  used  with  underwater  techniques.    It  is  recommended  that  the  handle  is  held  by  the 
operator without direct contact to the water, to avoid exposure. 
 
The Omnisound

®

 3000C Pro transducer contains a complete workstation which allows you to set output  directly  on 

the  transducer  handle.    The  power  up  and  down  arrows  allow  setting  of  the  power  or  intensity  of  the  ultrasound 
beam.  The timer starts when the power up arrow is released from a zero output starting position.  The timer stops 
(holds) when the output is reset to zero with the power down arrow.  This can be used if you are interrupted during 
treatment and you wish to maintain initial treatment time.  A continuous tone will be heard while adjusting output 
power up or down. 

NOTE:

  Output power is automatically decreased to zero upon frequency change.

 

 

 

 

 

 
 

OMNISOUND

®

 3000C PRO TRANSDUCER

Содержание 1203000C

Страница 1: ......

Страница 2: ...not be used for any other purpose Any additional copies of the Manual shall be ordered from ACP No changes or modifications shall be made to the Manual without prior review and written authorization...

Страница 3: ......

Страница 4: ...Ultrasound Pulsed Shortwave Diathermy devices and advanced Therapeutic Exercise Systems Our MEGAPULSE NEUROPROBE OMNISTIM OMNISOUND OMNIBAND OMNITEST OMNICYCLE and OMNIVR represent the most recent wor...

Страница 5: ...ocuments Caution electrical precautions Remote Control connection CE mark of confidence compliant to MDD 93 42 EEC Connection Electrode Cable IP2X Classification against intrusion of water dust accide...

Страница 6: ...CIPLES OF USE 15 Definitions 15 Universal Precautions Body Substance Isolation 15 Cleaning Disinfecting of the Omnisound 3000C Pro 15 Cleaning and Low Level Disinfection 15 Intermediate Level Disinfec...

Страница 7: ...crylate or high density polyethylene Arthroplasties the effect on bony ingrowth arthroplasties is not well defined for this reason the most prudent course is avoiding ultrasonic therapy over these are...

Страница 8: ...or neck triangle including stellate ganglion vagus nerve or laryngeal or pharyngeal muscle Particular care should be taken for patients with a known sensitivity to the carotid sinus reflex as carotid...

Страница 9: ...ontinue the treatment and shorten the treatment time on the next treatment session or use an alternative type of therapy or applicator placement Do not apply treatment directly over under hot or cold...

Страница 10: ...at make the Omnisound 3000C Pro truly the future of ultrasound therapy Controlled Depth of Penetration The Omnisound 3000C Pro enables you to treat soft tissue injuries superficially with 3 MHz and in...

Страница 11: ...FACTOR buttons are used to select the ultrasound output Continuous 100 for thermal applications Pulsed 5 10 20 25 33 or 50 for sub thermal applications NOTE Duty factor starts at continuous and goes...

Страница 12: ...display in watts power or watts cm intensity Q The cooling fan provides airflow through the unit when the unit is operational R The ELECTROSTIM INPUT PORT is used to connect the shrouded pin to banan...

Страница 13: ...off the system and connect the transducer to the Main output connector If both transducers are connected the main transducer will be selected 2 Omnisound 3000C Pro transducers contain their calibrati...

Страница 14: ...rapidly if the output is switched ON and the transducer is not in contact with the skin Power Supply Operation Omnisound 3000C Pro is a line power only device It is not intended for use with batteries...

Страница 15: ...ucers are fully waterproof and may be used with underwater techniques It is recommended that the handle is held by the operator without direct contact to the water to avoid exposure The Omnisound 3000...

Страница 16: ...t a lower output is recommended Avoid using maximum power levels For example a typical comfortable output should be 0 5 1 0 W cm2 4 Apply coupling medium plentifully to the area to be treated As the s...

Страница 17: ...s the active electrode Attach the self adhesive electrode to the patient s body Take the negative black tip pin and insert it into the shrouded pin to banana adapter Item No 2503036 Insert the adapter...

Страница 18: ...inal secretions is likely to occur Health care workers with exudating lesions or weeping dermatitis should refrain from all direct patient care and from handling patient care equipment until the condi...

Страница 19: ...creased risk of patient cross contamination After each use clean common contact surfaces such as control panel and ultrasound transducer with ACP germicidal wipes With a second ACP germicidal cloth wi...

Страница 20: ...e Omnisound 3000C Prounit 4 Select and cut with clean scissors a 2 foot length of plastic sleeve and slide it over the ultrasound handle and cable 5 Apply ultrasound gel into plastic transducer cover...

Страница 21: ...ing in pulsed or continuous modes at 1 0 MHz or 3 0 MHz for therapeutic applications Theory of Operation Ultrasonic therapy devices generally consist of a generator and a transducer The generator prod...

Страница 22: ...ted that the effective radiating area does not generally cover the entire surface area of the transducer effective radiating surface and varies with the crystal frequency when operated at different cr...

Страница 23: ...20 OMNISOUND 3000C PRO USER MANUAL COPYRIGHT 2001 2017 ACCELERATED CARE PLUS CORP ALL RIGHTS RESERVED...

Страница 24: ...rally produce beams of perfectly uniform intensity over the surface of the transducer Thus the measurement represents an average for the effective radiating surface area Beam Divergence The beam diver...

Страница 25: ...e BNR was 7 1 the maximum intensity in the tissue would equal 7 watts cm2 at certain portions of the field This should be kept in mind when determining output dosage It is also important to maintain c...

Страница 26: ...inuous overheating of the transducer Ultrasound Transmission and Absorption When ultrasonic energy is applied to the tissue it is absorbed in varying degrees This absorption of energy can increase tis...

Страница 27: ...supply adequate couplant apply the transducer to the patient and increase the power As you do so you will notice that the timer will change from 15 minutes to a lower value This will only occur if th...

Страница 28: ...ity contact the Service Center TRANSDUCER NOT PRESENT message is displayed on the unit Transducer may not be connected properly Check if transducer is plugged in correctly If not plug in transducer an...

Страница 29: ...andard 21 CFR 1050 10 instituted by the Food and Drug Administration Periodic calibration of treatment parameters having a direct influence on the physiological effect of ultrasound on tissue i e ultr...

Страница 30: ...n Rate 20 20 1 1 100 continuous continuous 2 1 50 2 ms 250Hz 3 1 33 2 ms 165Hz 4 1 25 2 ms 125Hz 5 1 20 2 ms 100Hz 10 1 10 2 ms 50Hz 20 1 5 2 ms 25Hz ULTRASOUND APPLICATOR Piezoelectric Disk Lead zirc...

Страница 31: ...t of the beam at 5mm from the transducer surface POWER AND INTENSITY Output Power 2cm2 5 cm2 1MHz 0 3W 0 12W Nominal 3MHz 0 3W 0 9W Nominal Output Intensity 1MHz 0 2W cm2 0 3W cm2 Nominal 3MHz 0 2W cm...

Страница 32: ...T 2001 2017 ACCELERATED CARE PLUS CORP ALL RIGHTS RESERVED ACP ULTRASONIC THERAPY APPLICATOR INTENSITY IMAGE AT 5mm FROM TRANSDUCER SURFACE Model Number 3000C Pro Serial Number 2945 Frequency 1 or 3 M...

Страница 33: ...3000C OMNISOUND 3000C PRO ULTRASOUND THERAPY SYSTEM 1 MHz and 3 MHz low BNR Ultrasound with Delta T dose control Shipping Weight 25 lbs 11 kg Standard Accessories ITEM ITEM NO DESCRIPTION 19856 AC Lin...

Страница 34: ...ces Infection Control 6 x 9 perforated Sheets 1200 sheets roll 50593 Barrier Film for Surfaces Infection Control 12 x 14 perforated Sheets 800 sheets roll 63574 Barrier Tubing 3 x 004 High Clarity 1 2...

Страница 35: ...ollowing have occurred 1 The product contains repairs or replacement parts not furnished by ACP 2 The product is damaged resulting from misuse or negligence 3 The product has been tampered with and or...

Страница 36: ...warranty ACP is not responsible for any loss or damage to the Equipment prior to receipt at the ACP Service Center Equipment returned for warranty service must be shipped complete with all accessorie...