background image

page 7

page 10

4.

External Switch Display Freeze

By connecting #10 and #12
of the communications port,
the tester instantaneously
captures the critical reading
and holds the display from
remote locations (use contact
closure and 

DO NOT

apply

voltage across #10 and #12).

5.

External Switch Display Clear

By connecting #8 and #12 of the communications port, dis-
play can be cleared from remote locations (use contact clo-
sure and 

DO NOT

apply voltage across).

(1)  Pay extra attention to avoid overload as display

value will not change during display hold.

(2)  Use contact closure only and 

DO NOT

apply

voltage across #10 and #12 port pins.

Use contact closure only and 

DO NOT

apply voltage

across #8 and #12 port pins.

2.

Mitutoyo Digimatic Signal

Connect the CB-304 cable to the communications port and
the device receiving the data. Set up parameters as instructed
from the Mitutoyo processor manual.

3.

±2 VDC Analog Signal

Connect the CB-104 analog cable to the communications port
and the device receiving the data.

Downloading Data from Memory

Choose between the following download methods.

Digimatic Data Download from Memory

1. Connect the tester and device receiving data with CB-304 cable. 

2. Turn on the tester. Press 

and 

for 3 seconds to enter

Power-On programming (CF9 flashes with solid F0).

3. Press 

to transmit all data.

4. Press 

again to exit.

RS-232C Data Download from Memory

Connect the tester and device receiving data with a CB-204 cable.
Use the I[CR] ASCII command to transmit data (uppercase ASCII
character format).

Recalling Data from Memory 

1. Turn on the tester.

2. Press 

and 

for 3 seconds to enter Power-On program-

ming (CF9 flashes with solid F0). Press 

and the display

cycles memory location and value. Press 

to increase loca-

tion and 

to decrease. Press 

to exit.

Clearing Data from Memory

1. Turn off the tester. 

2. Press 

again while holding 

to enter memory mode.

Single Memory Clear

A memory location with a dot at both ends is the last stored value
and the only one that can be erased. Press 

to erase and ErASEd

is displayed.  If you erase any other location Error is displayed. 

All Memory Clear

While a memory location or value is displayed, press 

for 3

seconds, all data is erased, ErASEd is displayed, then 

– – – – – 

and

flashing CF9 with solid End. Press to 

exit. 

(See page 14-15).

Содержание DTX2

Страница 1: ... authorization number and return the defective product to us with freight prepaid The foregoing warranty constitutes the SOLE AND EXCLUSIVE WARRANTY and we hereby disclaim all other warranties express statutory or implied applicable to the products and or software including but not limited to all implied warranties of merchantability fitness non infringement results accuracy security and freedom f...

Страница 2: ...its between lb in kg cm and N cm Programmable Setpoint LED s When high low setpoints are set LED indicates below NG within OK or above set point value NG CW icon Displayed for CW measurements CCW icon Displayed for CCW measurements Battery icon Flashes when the NI MH cells need to be charged PEAK icon When continuously displayed peak function is activated HOLD icon Displays when external hold sign...

Страница 3: ...se inter nal Ni MH batteries without plugging into AC power Turn on the tester and check that the display is at zero and overload indicator is not flashing PRECAUTIONS DTX2 Cap Torque Tester is a sensitive instrument The load cell can be damaged regardless of whether the unit is ON or OFF Follow these precautions to keep your DTX2 from being damaged 1 Do not exceed unit s capacity regardless if po...

Страница 4: ...apid torque changes over time Larger numbers allow more dampening Press SEND to select CF9End displays press SEND to exit Torque dampening rate displays each time the tester is turned on CF9 m2 indicator Default SC OFF Press PEAK or ZERO to cycle SC OFF CW and CCW or SC On CW and CCW Press SEND to select CF9End displays press SEND to exit Power Off programming clear data from memory Turn off the t...

Страница 5: ...LO displays then the low set value i e L 5 0 Press PEAK to increase and ZERO to decrease press SEND to select CF9 End displays Press SEND again to exit CF9 F2 Peak mode Default Or Press SEND Or or And displays Press PEAK or ZERO to cycle Press SEND to select Or Peak records the Peak in either CW or CCW during test And Peak records both the CW peak and CCW peak during a test Zero reset memory store...

Страница 6: ...example 5 lbf in is set as Low and 10 lbf in as High the ORANGE LED light for measurements less than 5 lbf in Low set point GREEN lights between 5 10 lbf in and RED lights over 10 lbf in High setpoint Setpoint output is available through the Communications port see page 7 After measuring press the button to transmit data to RS 232 models RS 232C or Digimatic devices Storing Data into Memory During...

Страница 7: ... Choose between the following download methods Digimatic Data Download from Memory 1 Connect the tester and device receiving data with CB 304 cable 2 Turn on the tester Press and for 3 seconds to enter Power On programming CF9 flashes with solid F0 3 Press to transmit all data 4 Press again to exit RS 232C Data Download from Memory Connect the tester and device receiving data with a CB 204 cable U...

Страница 8: ...arity Baud Rate 19200 bps RS 232C INTERFACE FUNCTIONS Upper case ASCII format COMMAND FUNCTION RESPONSE T CR Select real time mode R CR P CR Select peak mode R CR If OR peak is programmed P CR peak If AND peak is programmed P CR 1st time peak P CR 2nd time peak Z CR Tare Display R CR D CR Transmit display data direction value units mode go nogo overload CR direction CW CCW value 4 digits w decimal...

Отзывы: