DIMENSIONS

Model

Capacity (Resolution)

lbf-in

kgf-cm

N-cm

DTX2-15

15.00 (0.01 lbf-in)

20.00 (0.01 kgf-cm)

200.0 (0.1 N-cm)

DTX2-40

40.00 (0.01 lbf-in)

50.00 (0.01 kgf-cm)

500.0 (0.1 N-cm)

DTX2-85

85.0 (0.1 lbf-in)

100.0 (0.1 kgf-cm)

1000 (1 N-cm)

INTRODUCTION

Model DTX2 is a highly sophisitcated, cap torque tester which offers
programmable high/low setpoints for go/no go testing.  Store up to
1,000 values into memory, which can be transmitted using Digimatic
or RS-232 formats.  Use the Real Time mode to display torque tran-
sients, or the Peak mode to capture the peak torque achieved during
a test.  Select measuring units between lb-in, kg-cm and N-cm.

Programmable Setpoint LED’s

When high-low setpoints are set, LED
indicates below (-NG), within (OK),
or above set point value (+NG).

CW icon

Displayed for CW measurements.

CCW icon

Displayed for CCW measurements.

Battery icon

Flashes when the NI-MH cells need
to be charged.

PEAK icon

When continuously displayed, peak
function is activated. 

HOLD icon

Displays when external hold signal
is activated or SEND button is
pressed.

Torque Units

Displays selected measuring units.
(Lbf-in, kgf-cm and N-cm)

Overload Indicator

Flashes at 110% of rated capacity.

Alarm Icon

Auto Power Off icon

10

9

8

7

6

5

4

3

2

1

(A) CLAMP HANDLE

(E) AC ADAPTER/CHARGER RECEPTACLE

(ON BACK)

(B) 1" TALL HOLDING PEGS (CUSTOM

HEIGHT PEGS ARE AVAILABLE)

(C) SAMPLE TABLE

(D) COMMUNICATIONS PORT

RS-232C, Digimatic & ± 2VDC Analog

1

2

3

4

5

6

7

8

9 10

Содержание DTX2

Страница 1: ... authorization number and return the defective product to us with freight prepaid The foregoing warranty constitutes the SOLE AND EXCLUSIVE WARRANTY and we hereby disclaim all other warranties express statutory or implied applicable to the products and or software including but not limited to all implied warranties of merchantability fitness non infringement results accuracy security and freedom f...

Страница 2: ...its between lb in kg cm and N cm Programmable Setpoint LED s When high low setpoints are set LED indicates below NG within OK or above set point value NG CW icon Displayed for CW measurements CCW icon Displayed for CCW measurements Battery icon Flashes when the NI MH cells need to be charged PEAK icon When continuously displayed peak function is activated HOLD icon Displays when external hold sign...

Страница 3: ...se inter nal Ni MH batteries without plugging into AC power Turn on the tester and check that the display is at zero and overload indicator is not flashing PRECAUTIONS DTX2 Cap Torque Tester is a sensitive instrument The load cell can be damaged regardless of whether the unit is ON or OFF Follow these precautions to keep your DTX2 from being damaged 1 Do not exceed unit s capacity regardless if po...

Страница 4: ...apid torque changes over time Larger numbers allow more dampening Press SEND to select CF9End displays press SEND to exit Torque dampening rate displays each time the tester is turned on CF9 m2 indicator Default SC OFF Press PEAK or ZERO to cycle SC OFF CW and CCW or SC On CW and CCW Press SEND to select CF9End displays press SEND to exit Power Off programming clear data from memory Turn off the t...

Страница 5: ...LO displays then the low set value i e L 5 0 Press PEAK to increase and ZERO to decrease press SEND to select CF9 End displays Press SEND again to exit CF9 F2 Peak mode Default Or Press SEND Or or And displays Press PEAK or ZERO to cycle Press SEND to select Or Peak records the Peak in either CW or CCW during test And Peak records both the CW peak and CCW peak during a test Zero reset memory store...

Страница 6: ...example 5 lbf in is set as Low and 10 lbf in as High the ORANGE LED light for measurements less than 5 lbf in Low set point GREEN lights between 5 10 lbf in and RED lights over 10 lbf in High setpoint Setpoint output is available through the Communications port see page 7 After measuring press the button to transmit data to RS 232 models RS 232C or Digimatic devices Storing Data into Memory During...

Страница 7: ... Choose between the following download methods Digimatic Data Download from Memory 1 Connect the tester and device receiving data with CB 304 cable 2 Turn on the tester Press and for 3 seconds to enter Power On programming CF9 flashes with solid F0 3 Press to transmit all data 4 Press again to exit RS 232C Data Download from Memory Connect the tester and device receiving data with a CB 204 cable U...

Страница 8: ...arity Baud Rate 19200 bps RS 232C INTERFACE FUNCTIONS Upper case ASCII format COMMAND FUNCTION RESPONSE T CR Select real time mode R CR P CR Select peak mode R CR If OR peak is programmed P CR peak If AND peak is programmed P CR 1st time peak P CR 2nd time peak Z CR Tare Display R CR D CR Transmit display data direction value units mode go nogo overload CR direction CW CCW value 4 digits w decimal...

Отзывы: