ElektroPhysik
E-45
22.1 Measuring uncertainty
*
The specifications are based on calibration of GalvanoTest 3000
using ElektroPhysik coating thickness standards.
Rubber gasket /
mask
Diameter
Measuring
area
Measuring
uncertainty* in
percent of reading
Rubber gasket B
2,26 mm
4 mm
2
6%
Mask C
1,13 mm
1 mm
2
9%
Mask D
0,56 mm
0,25 mm
2
20%