E-42

ElektroPhysik

17.  Maintenance

Please keep the instrument clean. Avoid spilling solution
on the instrument or on the accessories.

The inner and outer measuring cells and power leads
must be kept clean to ensure error free operation.

Before filling in fresh solution, rinse the measuring cell
and rubber gaskets using distilled water. If the same
solution is used for the next measurement, it is sufficient
to clean the bottom side of rubber gasket after each
measurement.

The outer measuring cell may build up residue or metal
deposits after several measurements. This generally
does not affect measurement. It will be sufficient to rinse
the measuring cell with distilled water.

In case of very strong pollution or routinely once a month,
the outer measuring celll should be cleaned in a solu-
tion of 4 parts concentrated sulphuric acid (Caution:
caustic) and one part concentrated nitric acid. They
should be immersed in the solution for not longer than
10 to 15 seconds and then rinsed under running water.
The procedure can be repeated if necessary until the
components are clean and shiny after which they should
be tried and polished.

4.

Instrument does not switch off, no analogue meter
deflection:

-

unsuitable solution

-

 Coating has been removed within the first three
 seconds, i.e. de-plating speed is too high.

5.

Instrument does not switchoff, but clear meter
deflection:

-

sensitivity too high

6.

Coating not properly removed:

-

sensitivity too low

7.

Wrong measuring result:

-

check steps  2., 3. and 6. (section 16)

-

instrument misadjusted, no or wrong calibration

-

unsuitable solution

8.

On switch on the instrument does not show any
setting, statistics, date and time are deleted:

-

Lithium buffer battery discharge.

Please send the instrument to ElektroPhysik.

If the problem cannot be solved by means of above
steps, please contact your dealer or ElektroPhysik
customer support.

Содержание GalvanoTest 3000

Страница 1: ...stributed by ABQ Industrial LP USA Tel 1 281 516 9292 888 275 5772 eFax 1 866 234 0451 Web https www abqindustrial net E mail info abqindustrial net Distributed by ABQ Industrial LP USA Tel 1 281 516...

Страница 2: ...ries E 11 4 Taking Readings E 12 4 1 Adjusting the instrument E 12 4 1 1 Metal coating material METAL E 12 4 1 2 Measuring area AREA E 12 4 1 3 Deplating speed control SPEED E 13 4 1 4 Sensitivity SEN...

Страница 3: ...8 Special applications E 29 8 1 Convex surfaces E 29 8 2 Concave surfaces E 30 8 3 Gold coatings Au E 30 8 4 Multi layer coatings E 30 8 5 Chromed zinc coatings E 31 8 6 Diffusion layers E 32 8 7 Cur...

Страница 4: ...Print out of readings and statistics E 38 12 4 Deleting a previous reading E 38 12 5 Deleting a measuring series incl statistics 12 6 Exceeding storage capacity E 39 13 Date and Time E 39 14 Language...

Страница 5: ...on 1 1 Data processing unit Front view 4 000 START STOP SET PRINT STATSCLEAR VOLTAGE THICKN GalvanoTest 3000 Physik Ni B 5 4 mm M E T A L A R E A S P E E D S E N S C A L U N I T 1 2 3 4 5 6 7 8 9 17 1...

Страница 6: ...lektroPhysik 1 2 Data Processing Unit Back view GalvanoTest CE Ger te N Typ Version 1 00 G Test 2000 3020001 G Test 3000 3040001 ElektroPhysik K ln Made in Germany 19 20 24 23 22 21 Sicherungs halter...

Страница 7: ...plating speed sensitivity calibration thickness of coating thickness standard measuring unit date and time language 6 SET key for setting parameters selected by ARROW keys 7 START STOP key for manual...

Страница 8: ...60Hz Switching from 220V to 110V Remove fuse holder 24 turn 180 degrees and put in place again 10 CLEAR key for deleting the last reading setting default de plating speed default sensitivity and defa...

Страница 9: ...g the inner measuring cell 32 circulation pump to ensure a steady circulation of the electrolyte solution during measurement 33 adjusting screw to adjust circulation Important note Please do not chang...

Страница 10: ...E 6 ElektroPhysik Fig 3a 34 38 36 37 35 31a 30 31 33 32 Suction hose to be connected to the circulation pump 32 Suction pipe...

Страница 11: ...drain of solution 44 inner measuring cell type II to be filled with the electrolyte solution diagonal drain of solution see Electrolyte solution selection table 44a Max Min marking limit for electrol...

Страница 12: ...el measuring cell is placed on the sample to be measured The cell is filled with an electrolyte solution which is selected according to the coating base combination Arubber gasket placed between the c...

Страница 13: ...mm 2 26 mm mask Nr C 1 00 mm 1 13 mm mask Nr D 0 25 mm 0 56 mm electrolyte cup K for small parts and wires The de plating speed is adjustable in a wide range from 0 3 to 40 m min The readings may be...

Страница 14: ...be linked to an IBM PC or any compatible computer or printer 3 Supply schedule 3 1 Standard supply GalvanoTest 3000 processing unit interface for MiniPrint 4100 data printer or PC socket for Y T write...

Страница 15: ...es 100 ml each coating thickness standards s also section 4 2 3 index ST electrolyte selection spare rubber pencil centering device Z2 spare pipette spare rubber gaskets B 4 00 mm masks C 1 00 mm mask...

Страница 16: ...to METAL 16 by pressing SET key Use the ARROW keys for setting your metal selection The options are shown in the following order Cr Ni Cu Ms brass Zn Ag Sn Pb Cd Au A set of further 8 metal memories M...

Страница 17: ...mils are displayed on the alpha numerical display To reset to default de plating speed press CLEAR key 10 This default value depends on the metal selection Note The de plating speed also depends on t...

Страница 18: ...est 3000 Physik Ni B 5 4 mm M E TA L A R E A S P E E D S E N S C A L U N IT If the required sensitivity is different from the default sen sitivity an asterix will be displayed above the cursor For mor...

Страница 19: ...holder can be used section 6 Place an apporpriate rubber gasket 45 onto the outer measuring cell 42 i e rubber gasket B 4 mm or the rubber gasket with 1 5mm if you use the mask C 1 mm or D 0 25 mm No...

Страница 20: ...b L e a d Tin Tin S n Tin Zinc4 Zinc Zn N o n m e ta l E 4 E 1 8 E 5 E 11 E 4 E 4 E 1 4 E 1 7 8 E 1 5 8 E 4 8 E 7 E 7 8 E 6 S ilve r E 1 5 8 A lum inium E 4 8 E 5 8 E 11 8 E 4 8 E 4 8 E 1 4 8 E 1 7 8...

Страница 21: ...ored The reading display 2 continuously shows the coating thickness being removed During measuring it can be switched from thickness to cell voltage by pressing THICKNESS VOLTAGE key 8 with the corres...

Страница 22: ...Clean with a soft paper towel Then proceed according to section 4 3 If it is necessary to change the electrolyte solution e g for measuring another coating substrate combination or if the electrolyte...

Страница 23: ...ed as soon as a volume of 300 m mm chrome has been de plated 4 4 2 Y T writer output During measurement the cell voltage can be taken at the Y Twriter output 21 and displayed in the form of a voltage...

Страница 24: ...y obtained by pressing THICKNESS VOLTAGE 8 key Keep on pressing more than one second after measurement If you press THICKNESS VOLTAGE key during meas urement the measuring current will continue to flo...

Страница 25: ...mplete print out by pressing PRINT key 9 4 5 2 Readings and statistics If the MiniPrint 4100 data printer is connected during measurement the readings are printed out after each measurement The statis...

Страница 26: ...e for pin point positioning of the measuring cell when using the C mask 1 00 mm or D mask 0 25mm 56 bottle holder for holding up to 7 electrolyte bottles 100 ml 57 positioning arm serves as stopper fo...

Страница 27: ...e centering device lights up Place the measuring cell upon the sample see section 4 2 2 and remove the centering device For more information on the use of the centering device in connection with the m...

Страница 28: ...available as an option 60 rubber protection jaws 61 ball joint for positioning the test sample 62 suction plate for ensuring stable position of the small parts holder during measuring Rubber protectio...

Страница 29: ...6 instead of the rubber gasket holder 42c see Fig 4 Make further adjustments according to section 4 Dip the sample into the electrolyte cup ensuring that the area dipped into the cup approximately cor...

Страница 30: ...ET key The instrument now calculates displays and stores the thickness value of the coating Note Depending on the type and thickness of coating wires 0 3 mm may also be measured in the electrolyte cup...

Страница 31: ...ea as follows F d x x l F Area d Diameter of wire 3 14 l length of de plated area Use ARROW keys for setting the actual measuring area e g 5 36 mm2 Press SET key to show the reading as calculated by m...

Страница 32: ...to insulate the end of wire completely If necessary use lacquer Dip the wire into the electrolyte cup ensuring that the punched hole of the C mask is completely immersed into the solution Continue as...

Страница 33: ...tion with the C or D mask Use abrasive paper to adapt the circular gasket to the shape of sample Samples 4mm e g wires electrolyte cup Wires 2 mm electrolyte cup and mask The rubber gasket B should no...

Страница 34: ...ing the modified rubber gasket D Au on the measuring stand full contact pressure must be applied to ensure maximum sealing of the mask on the measuring surface The D and C masks are not suitable for m...

Страница 35: ...a te bottle Rinse the inner and outer measruing cells twice using distilled water Move the cursor under METAL on the alphanumeric display 3 to adjust the next mate rial to be measured Fill fresh elect...

Страница 36: ...wer voltage value For setting optimum sensitivity make sure the instrument does not switch off before the second voltage jump see also section 10 Sensitivity The print out of voltage curve on MiniPrin...

Страница 37: ...ck coatings the instrument may switch of prematurely i e before the coating has been removed completely In such cases 0 00 will be shown on display and no reading will be stored 8 8 Nickel on copper o...

Страница 38: ...asurement series however cannot be stored until previous readings in the statistics have been deleted see section 12 5 Deleting a measuring series incl statistics 9 1 Calibrating by means of a coating...

Страница 39: ...ctor If coating thickness standards are not available calibration can also be performed by means of a calibration factor providing instrument settings are known Calibration by means of a factor is onl...

Страница 40: ...can be observed on the de plating area To ensure a smooth and even de plating area it is recommended to increase the test voltage E g for nickel it is recommeded to increase the test voltage from 1 0...

Страница 41: ...due to diffe rent de plating speeds We recommend to calibrate at the same de plating speed as set for later measurement Impurities or alloy components in the coating material may lead to erratic read...

Страница 42: ...eadings will be printed out printer ON The statistical values can be displayed or printed out as often as required and at any time after measurement 12 3 Print out of readings and statistics Press Gal...

Страница 43: ...crystal controlled time clock which includes an integrated leap year control up to the year 2100 The data printer MiniPrint 4100 pro vides print outs including current date and time Date and time can...

Страница 44: ...nguage is shown Your selection is valid without need to confirm 15 Error messages Error messages appear on the alphanumerical display Maximum Calibration CAL sensitivity SENS de plating speed SPEED or...

Страница 45: ...atistics Error message when trying to call or print any statistics but no data are available in the statistics memory 16 Further sources of error Apart from the error sources as described under sectio...

Страница 46: ...uric acid Caution caustic and one part concentrated nitric acid They should be immersed in the solution for not longer than 10 to 15 seconds and then rinsed under running water The procedure can be re...

Страница 47: ...niPrint 4100 press printer ON OFF key for switching the printer on 20 After sales service Please return defective or faulty instruments to your local dealer or to ElektroPhysik for checking or repair...

Страница 48: ...ble Display of readings 4 digits digital display with floating point Alphanumerical display 16 digits display of instrument settings and operating steps Analogue instrument for monitoring the de plati...

Страница 49: ...s are based on calibration of GalvanoTest 3000 using ElektroPhysik coating thickness standards Rubber gasket mask Diameter Measuring area Measuring uncertainty in percent of reading Rubber gasket B 2...

Страница 50: ...11 E 25 E 28 convex surfaces E 29 copper E 8 E 30 E 33 curvature radius E 41 D D mask E 22 E 29 data printer E 3 E 4 E 9 E 10 E 11 E 20 E 21 E 39 date and time E 3 E 9 E 38 E 39 E 42 de plating speed...

Страница 51: ...ssivation layers E 31 positioning arm E 11 E 22 E 23 power leads E 42 PRINT key E 3 E 21 E 38 E 43 R reading display E 12 E 13 E 14 E 17 E 35 E 36 E 41 rubber gasket E 7 E 8 E 9 E 11 E 15 E 23 E 25 E...

Страница 52: ...E 16 zinc E 8 E 9 E 15 E 31 E 33 Distributed by ABQ Industrial LP USA Tel 1 281 516 9292 888 275 5772 eFax 1 866 234 0451 Web https www abqindustrial net E mail info abqindustrial net Distributed by A...

Отзывы: