![Abilize AETHOS SLE Скачать руководство пользователя страница 18](http://html1.mh-extra.com/html/abilize/aethos-sle/aethos-sle_user-manual_3392425018.webp)
P. 18
7. TROUBLESHOOTING
There is a self-diagnostic warning light on the
control panel. To check the self-diagnostic
warning light, turn on the power and count
the number of times the warning light
flashes.
Self-Diagnostic
Warning Light
Normal: Solid light
Abnormal: Flashing light
FLASH DESCRIPTION
MEANING
1
Low Charge
The batteries are running low.
Recharge the batteries for a minimum of 12 hours .
2
Low Battery Voltage
The batteries have run out of charge.
Recharge the batteries .
3
High Battery Voltage
Battery voltage is too high. This may occur if overcharged &/or
traveling down a long slope.
If traveling down a slope, reduce your speed to minimize the amount of
regenerative charging.
Check the battery and associated connections and wiring.
4
Current Limit Time-Out
The motor has been exceeding its maximum current rating for too
long. This may be due to a faulty motor.
Check the motor and associated connections and wiring.
Turn the controller off, leave for a few minutes and turn back on again.
5
Park Brake Fluid
Either a park brake release switch is active or the park brake is
faulty.
Check the park brake and associated connections and wiring.
Ensure any associated switches are in their correct positions.
6
Throttle OONAPU
The Throttle is out of neutral when turning the controller on.
Ensure the throttle is in neutral when turning the controller on.
The Throttle may require re-calibration .
7
Speed Pot Fault
The throttle, speed limit pot or their associated wiring may be
faulty.
Check the throttle and speed pot and associated connections and wiring.
8
Motor Fault
The motor or its associated wiring is faulty.
Check the motor and associated connections and wiring.
9
Other Error
The controller may have an internal fault.
Check all connections & wiring .