9
3.1.1. Adjust the tonic-clonic seizure notification delay
Setting the time delay before the alarm is triggered.
When the device detects faster movement (between 3–20Hz), its internal timer switches on. The
device gives a notification if the movement continues for longer than the preset time. The delay can
be set to 10, 13, 16 or 20 seconds.
Choose the desired delay according to the following table. If there are problems with false alarm,
increase the time delay.
Setting the fast movements duration after
which the notification goes off
Switch 1
Switch 2
10 seconds
ON (up)
ON (up)
13 seconds (default)
OFF (down)
OFF (down)
16 seconds
ON (up)
OFF (down)
20 seconds
OFF (down)
ON (up)
3.1.2. Adjust the Bed-Exit Notification Delay
The device can be set to trigger a notification when a person leaves the bed and the system does
not notice any micro-movements.
Time delay
Switch 3
Switch 4
Switch 5
No delay
OFF (down)
OFF (down)
OFF (down)
Disabled (default)
ON (up)
OFF (down)
OFF (down)
3 min Delay
ON (up)
ON (up)
OFF (down)
6 min Delay
OFF (down)
OFF (down)
ON (up)
10 min Delay
ON (up)
OFF (down)
ON (up)
15 min Delay
OFF (down)
ON (up)
ON (up)
30 min Delay
ON (up)
ON (up)
ON (up)
3.1.3. Activate or deactivate the SW1 switch function
SW 1 switch function
Switch 6
The SW1 switch acts as the On/Off switch which means that the device
can be switched on or off by pressing the switch for three (3) seconds.
OFF (down)
The SW1 switch does not act as the On/Off switch. The device is always
on when it is connected to the power supply or when the batteries are in.
ON (up)
3.1.4. Adjust the notification sound volume
NOTE:
Adjusting the volume may lead to the notification sound not being heard.
The sound signal must only be used when care persons are near the control unit. The volume can
be set to loud, normal, quiet, or mute.
The sound signal stops when the reset button is pressed and when a seizure is over. If the user
leaves the bed, the sound signal stops when the user returns to the bed or when the SW1 switch is
pressed.
Level
Switch 7
Switch 8
Very loud
OFF (down)
OFF (down)
Normal
ON (up)
OFF (down)
Quiet
OFF (down)
ON (up)
Mute
ON (up)
ON (up)