16 

11.  Installation of the bed sensor 

Place the bed sensor across the bed, under the mattress at 
approximately chest height. (picture 12) 

To prevent the bed sensor from moving, fix the sensor to the 
bed bottom or mattress using the double sided tape provided 
or e.g. cable ties (picture 26). The tape will not leave a mark 
and it will come off as a whole if removed. 

Check at least once a week that the bed sensor is properly 
positioned. 

If using the bed sensor with a spring mattress, place the sensor 
beneath the mattress and on top of the box spring or 
flat/platform surface. Place above spring mattress and below a 
mattress pad if bed is especially springy, you do not have a flat 
surface to place sensor on, individual being monitored weighs 
less than 35 lbs/15 kgs, or if alarm does not trigger with sensor 
below mattress after increasing sensitivity to maximum level allowed (Manufacturer 
recommendation is no higher than 8). Thickness of mattress and weight of individual may 
require that sensor be placed above the mattress and below a mattress pad. 

Always place the bed sensor under the mattress or mattress topper, never just under the 
sheet. The bed sensor must not come into direct contact with a person! 

The bed sensor is designed to last for a minimum of two (2) years placed under a foam 
mattress and against the hard base of a bed frame. 

With a spring mattress, the service life of the bed sensor may be considerably shorter. 
When the bed sensor is placed on a spring mattress, the user's weight and movement may 
cause the sensor to crumple, which may affect the sensor's performance. The 
manufacturer recommends that users replace the sensor when it starts to look crumpled. 
When used with a spring mattress, the sensor should be replaced annually. The warranty 
does not cover damage caused by crumpling. 

If the bed sensor is used with a pressure care mattress that is adjusted using a 
compressor pump, the mattress may interfere with the sensor's performance. If you are 
unsure about the suitability of your mattress, please contact the manufacturer. 

 

 Keep device cords out of the reach of children (risk of strangulation). Use protective 

cord covers or cable ties to securely attach and hide cords to prevent strangulation. Keep 
cords as short as possible and secure them so that they are out of reach of children. 

 

 

 

Содержание 464000

Страница 1: ...onitor Art no 464000 464031 Note The device must not be used if an operational failure of the device might lead to a delay in getting the appropriate treatment or medication for the patient which coul...

Страница 2: ...2...

Страница 3: ...6 3 SW 1 switch function switch 6 12 6 4 Adjusting the notification sound volume switches 7 and 8 12 7 Inserting batteries and battery service life 13 8 5V AC External power supply optional 14 9 Conne...

Страница 4: ...al of the device after use 23 21 Declaration of conformity EU 23 22 Abilia limited warranty statement 23 22 1 Your guarantee 23 22 2 Terms 23 22 3 Exclusions and limitations 24 22 4 Your legal rights...

Страница 5: ...nications or local area networks If this device is used with a pressure care mattress filled using a compression pump the device may not function normally in some cases Equipment is not suitable for u...

Страница 6: ...se chapter Electromagnetic conditions Symbol of European Waste Electrical and Electronic Equipment Directive WEEE Directive 2002 96 EC on waste electrical and electronic equipment in these instruction...

Страница 7: ...certified in the external power supply Product is C TICK certified in the external power supply Product is China RoHS 30 certified in the external power supply Product is BSMI certified in the externa...

Страница 8: ...consists of a bed sensor located under the mattress and a control unit where the bed sensor is connected Control unit should be placed on a table clipped on bed or mounted on wall The control unit no...

Страница 9: ...operator is the caregiver of the user The operator needs to be adult who fully understands instructions and limitations of use mentioned in these instructions for use Otherwise no special skills or k...

Страница 10: ...details picture 2 In the event of a power failure two high quality 1 5V AA alkaline batteries can be used as an emergency power supply Do not use any rechargeable batteries or lithiumion batteries Re...

Страница 11: ...er supply and remove nurse call cable if used from connector X2 before opening the lid Remove the batteries and disconnect the power supply before setting up the DIP switches in order to implement the...

Страница 12: ...switched on or off by pressing the switch for three 3 seconds OFF down The SW1 switch does not act as the On Off switch The device is always on when it is connected to the power supply or when the bat...

Страница 13: ...is someone in bed 75 of that time there are two notifications per day and sound notification is on 30 seconds each time When the batteries are getting low the red LED will begin to flash slowly A beep...

Страница 14: ...s connected the batteries act as backup power supply in the event of a power failure All alkaline batteries self discharge and start to leak when empty contaminating the device Ensure that batteries a...

Страница 15: ...the bed sensor picture 13 the power supply delivered as an accessory picture 14 and any connector cables picture 15 according to the pictures 9 1 X2 AUX Connector Pin Order picture 16 From left to rig...

Страница 16: ...place the bed sensor under the mattress or mattress topper never just under the sheet The bed sensor must not come into direct contact with a person The bed sensor is designed to last for a minimum o...

Страница 17: ...press the switch for too long you might accidentally switch off the device The notification sound will also stop when the fast paced movements stop 12 3Bed exit alarm bypass switch The SW1 switch act...

Страница 18: ...f No one is on the bed 13 2Blue Device on standby Light is off Device is switched off Light flashes slowly Device is switched on Light flashes rapidly a few times Device triggers a notification 13 3Re...

Страница 19: ...re high priority signals that will overrule information signals if both signals activate simultaneously 14 1Alarm signals Two high and low BEEPS in a loop Fast movements alarm goes off Three high and...

Страница 20: ...bed should not last long enough for the notification to sound Person weight Switch position in parenthesis if false alarms occur 75 kg 1 0 50 75 kg 2 1 35 50 kg 3 2 or 1 25 35 kg 4 or 5 3 or 2 15 25 k...

Страница 21: ...of the bed sensor under the mattress The correct position is at the chest height of the user and across the bed 3 Power supply battery check if power supply is used Make sure that power supply and bed...

Страница 22: ...e device on only after the person has fallen asleep If the bed exit alarm is enabled DIP switch 3 in DOWN position and sensitivity is not set high enough alarm may trigger Either disable alarm DIP swi...

Страница 23: ...ntact your dealer distributor or manufacturer To avoid any unnecessary inconvenience on your part we recommend you read these instructions for use carefully before seeking guarantee service 22 1Your g...

Страница 24: ...beyond specifications or features described in the instructions for use or modifications to the product to conform it to national or local technical or safety standards in countries other than those...

Страница 25: ...xample some countries prohibit the exclusion or limitation of damages resulting from negligence gross negligence wilful misconduct deceit and similar acts Abilia s liability under this guarantee will...

Страница 26: ...eset switch 8 pcs DIP switches for settings volume delay one 10 position rotary switch for adjusting sensitivity Signal lights 3 LEDs green blue and red Delays Movement notification delay alternatives...

Страница 27: ...W 15 W Output voltage 5 VDC Output current 1 2 A 3 0 A Electrical safety class Class II 23 4Ambient conditions Operating temperature 10 40 C Storage and transport temperature 30 50 C Relative humidity...

Страница 28: ...pecified below The customer or the user of the equipment should assure that it is used in such an environment RF emissions CISPR 11 Group 1 The Emfit Tonic Clonic Seizure Monitor uses RF energy only f...

Страница 29: ...61000 4 5 1 kV line s to line s IEC 60601 1 2 test level Mains power quality should be that of a typical commercial or hospital environment Voltage dips short interruptions and voltage variations on...

Страница 30: ...each frequency range b Interference may occur in the vicinity of equipment marked with the following symbol NOTE 1 At 80 MHz and 800 MHz the higher frequency range applies NOTE 2 These guidelines may...

Страница 31: ...to frequency of transmitter 150 kHz 80 MHz d 1 2 P 80 MHz 800 MHz d 1 2 P 800 MHz 2 5 GHz d 1 2 P 0 01 0 12 m 0 12 m 0 23 m 0 1 0 38 m 0 38 m 0 73 m 1 1 2 m 1 2 m 2 3 m 10 3 8 m 3 8 m 7 3 m 100 12 m...

Страница 32: ...32 26 Appendix Related pictures...

Страница 33: ...33...

Страница 34: ...34...

Страница 35: ...35...

Страница 36: ...36 464000 464031_Manual Emfit Tonic Clonic Seizure Monitor_EN_RevF1 docx Dec 2017...

Отзывы: