6.2 Setting the Bed-Exit Notification Delay
The device can be set to trigger notification when person leaves the bed and system does
not notice any micro-movements.
Time delay
Switch 3
Switch 4
Switch 5
No Delay
OFF (down) OFF (down) OFF (down)
Disabled (default)
ON (up)
OFF (down) OFF (down)
3min Delay
ON (up)
ON (up)
OFF (down)
6min Delay
OFF (down) OFF (down) ON (up)
10min Delay
ON (up)
OFF (down) ON (up)
15min Delay
OFF (down) ON (up)
ON (up)
30min Delay
ON (up)
ON (up)
ON (up)
6.3 SW 1 switch function (switch 6)
SW 1 switch function
Switch 6
The SW1 switch acts as the On/Off switch which means that the
device can be switched on or off by pressing the switch for three
(3) seconds.
OFF
(down)
The SW1 switch does not act as the On/Off switch. The device is
always on when it is connected to the power supply or when the
batteries are in.
ON (up)
6.4 Adjusting the notification sound volume (switches 7 and 8)
There are four volume levels: Very loud, normal, quiet and mute. The factory setting is
very loud. Volume level should be set so that notification is audible in the environment of
use.
Level
Switch 7
Switch 8
Very loud
OFF (down) OFF (down)
Normal
ON (up)
OFF (down)
Quiet
OFF (down) ON (up)
Mute
ON (up)
ON (up)
The notification sound stops when the SW1 switch is pressed or when the seizure stops.
Remove the batteries and disconnect the power supply before setting up the DIP
switches in order to implement the new settings.