ABEM Terrameter SAS 1000 / SAS 4000
- 37 -
6.2.5
Using the LUND Imaging System Software
In the start-up menu select "LUND Imaging System" and press
. This starts the LUND
program as described in the following.
The first menu (after the welcome menu) is the
Record Manager. Select the preferred mode (SP,
Resistivity or IP).
╔════════════
RECORD MANAGER
════════════╗
║
Mode: Resistivity
║
║
Record:
║
║
║
║
║
║
║
║
║
╚
=Measure,
=Quit,
=Next/Prev Page
═╝
If you move to Record and press
you get the
options to create a new record, open an existing or
delete record(s).
╔════════════
RECORD MANAGER
════════════╗
║
Mode: Resistivity
╔═══
Record
═══╗
║
║
Record:
║
<New>
║
║
║
║
<Open>
║
║
║
║
<Delete>
║
║
║
║
<Delete All>
║
║
║
╚════════════╝
║
╚
=Measure,
=Quit,
=Next/Prev Page
═╝
If you select “New” you get the option to name a
new record. Either accept the default name, or
define a new. Use the
knob to change
characters, and the
/
knob to move between
fields. If you name the record “00005”, the record
name will become LR00005. LR means LUND
and Resistivity. If the mode was IP, the record
name would become LI00005.
╔════════════
RECORD MANAGER
════════════╗
║
Mode: Resistivity
║
║
Record:
╔════════════┌──────┐════╗
║
║
║
New record:
│
00005
│
║
║
║
╚════════════└──────┘════╝
║
║
║
║
║
╚
=Measure,
=Quit,
=Next/Prev Page
═╝
If you select a new record this menu appears. The
smallest electrode distance is simply the electrode
separation in meters. The power line frequency is
either 50 Hz or 60 Hz. The midpoint coordinates
can be given in units of electrode spacing. The
base
ref.
is
the
coordinates
and
the
azimuth/inclination of the origo of the base. To
continue, select OK and press
or simply press
.
╔══════════
(Res.) LR00005.S4K
══════════╗
║
Smallest electrode distance: 2.0 m
║
║
Power line freq.: 50 Hz
║
║
Midpoint X: 0.0 ( 0.0 m )
║
║
Y: 0.0 ( 0.0 m )
║
║
<Base Ref.>
║
║
<Ok> <Cancel>
║
╚══════════════════════════════════════╝
This brings up the menu where you select protocol
file(s). At the first entry you have to select "Add"
and press .
╔══════════════════════════════════════╗
║
1: 5:
║
║
2: 6:
║
║
3: 7:
║
║
4: 8:
║
║
║
║
<Add> <Clear> <Ok> <Cancel>
║
╚══════════════════════════════════════╝
Move to the "Select" entry and press .
╔════════════
Add protocol:
════════════╗
║
║
║
Select: POL2L
║
║
║
║
║
║
║
║
<Ok> <Cancel>
║
╚══════════════════════════════════════╝
This brings up a list of all possible protocol file. In
this list, move with the
to a suitable protocol,
e.g. Wenner_L. Then press , and you are back in
the menu shown above. Go to <Ok> and press
.
The message "Validating protocol ..." will appear
for a couple of seconds.
╔════════┌─────────────────┐═══════════╗
║
│
WENB_L
│
║
║
Select:
│
WENB_S
│
║
║
│
WENFASTL
│
║
║
│
WENFASTS
│
║
║
│
WENNER_L
│
║
║
<Ok>
│
WENNER_S
│
║
╚════════└─────────────────┘═══════════╝