ABEM Terrameter SAS 1000 / SAS 4000
- 96 -
14.4
WE WENNER_S: 1-CHANNEL WENNER CVES WITH
4 ELECTRODE CABLES
This combination of protocol files is intended for Wenner-
CVES roll-along measurements,
and has been found very useful for a variety of applications. Note that it is important to select
the protocol files in the correct order, starting with the long layout (WENNER_L) before
selecting the protocol for short layout (WENNER_S). It gives a dense near-surface cover, and
a slightly sparser measurement pattern at long electrode spacings. The WENNER_L protocol
supplied here has a higher data density than in previous releases, being identical to the
previously supplied WENNERXL.
WENNER_S is designed only to supplement WENNER_L for the shortest electrode spacings,
and is generally not suitable for use alone if only the two center cables are used. If measuring
with two cables only, either use WENNER_L and exclude all electrode positions on the outer
electrode cables, or use a protocol dedicated for using two electrode cables only (see below).
Protocol
Array Type Address File
Size .ORG Size .UP/.DWN
WENNER_L
Wenner-
LONG
258
113
WENNER_S
Wenner-
SHORT
87
50
Sum
345
163
Figure 28: Data cover of standard Wenner CVES using WENNER_L and WENNER_S
for roll-along with three stations (take-out spacing 1metre).