4
OVER
4 Perform a Control Solution Test
Purpose
The purpose of the control solution test is to:
• Ensure that your receiver and test strips are working properly prior to calibration
or blood glucose testing.
• Check that you are following the correct testing procedure.
• Practice Blood Glucose mode testing without having to use your own blood.
A Control Solution test is not the same as a calibration. See Section 7 for
calibrating the system.
Start
Do this procedure when you:
• Set up your receiver for the first time.
• Open and begin using a new vial of test strips.
• Suspect that your receiver or test strips are not working properly.
• Suspect that your glucose test results are not accurate.
• Suspect that your test results are not consistent with how you feel.
• Drop, damage, or expose the receiver to liquids.
• Are advised by your healthcare team to do so.
Materials
Gather these materials before you begin:
• FreeStyle Navigator Receiver.
• FreeStyle Control Solution.
• FreeStyle Test Strips.
Helpful Information
Caution:
•
Do NOT
use FreeStyle Control Solution for calibration. Do
NOT
perform the control solution test
when the system asks you to ‘Do BG Test’ for calibration.
• Results from Control Solution tests do
NOT
reflect your blood glucose level.
Important:
The control solution range is a target range for the control solution only. It is
NOT
a target
range for your blood glucose levels.
Important:
If your control solution test results continue to fall outside of the range printed on the test
strip vial, the receiver may not be working properly. Do
NOT
use the system to test your glucose levels.
Call Customer Care.
Ensure accurate control solution results by doing the following:
• Use
ONLY
FreeStyle Control Solution and FreeStyle Test Strips.
• Replace the cap on the control solution vial immediately after using.
• Do
NOT
use your control solution past the expiration date or the discard date.
• Do
NOT
add water or any liquid to the control solution.
• Perform control solution tests only between 59° and 104° F (15° and 40° C).
For important test strip information, including detailed storage and usage information, refer to the
FreeStyle Test Strip package insert.
Section 4
27
Perform a Control Solution Test
Содержание FreeStyle Navigator
Страница 1: ...User s Guide CAUTION Federal law restricts this device to sale by or on the order of a physician ...
Страница 2: ......
Страница 4: ......
Страница 8: ......
Страница 9: ...1 Getting Acquainted ...
Страница 23: ...2 Install or Replace Transmitter and Receiver Batteries ...
Страница 29: ...3 Set the Time and Date ...
Страница 34: ...Customer Care 1 866 597 5520 ...
Страница 35: ...4 Perform a Control Solution Test ...
Страница 42: ...Customer Care 1 866 597 5520 ...
Страница 43: ...5 Insert or Remove Your Sensor ...
Страница 54: ...Customer Care 1 866 597 5520 ...
Страница 55: ...6 Attach Your Transmitter ...
Страница 61: ...7 Calibrate Your System Test Blood Glucose Manually ...
Страница 74: ...Customer Care 1 866 597 5520 ...
Страница 75: ...8 Set Review or Change the Alarm Settings ...
Страница 89: ...9 Daily Use ...
Страница 96: ...Customer Care 1 866 597 5520 ...
Страница 97: ...10 Respond to Alarms Errors and Problems ...
Страница 119: ...11 Add Events ...
Страница 125: ...12 Review Reports and Edit Events ...
Страница 135: ...13 Clean Your System ...
Страница 136: ...Section 13 Key Terms n Receiver n Transmitter n Contact Points Customer Care 1 866 597 5520 ...
Страница 139: ...14 Link and Unlink Your Transmitter and Receiver ...
Страница 140: ...Section 14 Key Terms n BG Mode n CM Mode n Link Unlink n Receiver n Transmitter Customer Care 1 866 597 5520 ...
Страница 144: ...Customer Care 1 866 597 5520 ...
Страница 145: ...15 Reset User Settings ...
Страница 146: ...Customer Care 1 866 597 5520 ...
Страница 149: ...16 Change Country Settings ...
Страница 155: ...17 Appendix A Site Maintenance ...
Страница 158: ...Customer Care 1 866 597 5520 ...
Страница 159: ...18 Appendix B Specifications ...
Страница 173: ...19 Appendix C FCC Compliance Information ...
Страница 174: ...Customer Care 1 866 597 5520 164 Section 18 Appendix B Specifications ...
Страница 182: ...Customer Care 1 866 597 5520 ...
Страница 183: ...20 Glossary ...
Страница 184: ...Customer Care 1 866 597 5520 174 Section 19 Appendix C Classification Compliance Declaration ...
Страница 191: ...21 Index ...
Страница 192: ...Customer Care 1 866 597 5520 ...
Страница 195: ......