21
3
Compare the control solution result to the range printed on the test strip vial . The result
should fall within the range .
Notes:
•
If you see an error message on the screen, your meter will explain what it means,
its possible causes, and what to do next .
•
The result will be marked as a control solution test in your logbook . It will not be
included in your charts and reports .
•
You may also mark a control solution test after it is performed by following these
steps:
a . At the Home screen, tap Logbook
.
b . Tap your most recent test . You can only modify this result .
c . Tap the down arrow . Then tap add or edit notes .
d . Tap the option for control solution test . Then tap OK .
e . You can now see your most recent result marked as a control solution test .
Out of Range Control Solution Results
•
Repeat the test if control solution results are outside the range printed on the test
strip vial .
•
Stop using the meter if control solution results are consistently outside the range
printed on the test strip vial . Contact Customer Service .