Set AUDIO/VIBRATE.
29.
Note:
The
Audio/Vibrate
setting establishes whether your pump will notify you with
audible sounds or vibrations.
Notice
a.
Next is highlighted.
Scroll to highlight
b.
Pump.
Press
c.
Enter button.
Scroll to select
d.
Audio (default) or Vibrate.
Press
e.
Enter button.
If you select
Then do this
Audio
Notice Audio is highlighted.
Press Enter button.
Scroll to select Low, Medium, or High.
Press Enter button.
Verify AUDIO/VIBE
30.
setting.
Review your Pump setting on the screen.
If setting is
Then do this
Correct
Notice Next is highlighted.
Press Enter button.
Incorrect
Repeat Step 29.
Steps
Do This
Set OCCLUSION ALARM.
31.
Note:
The
Occlusion Alarm
alerts you that there is pressure in your pump or tubing
sensed the pump. This can result in insulin
NOT
being delivered to your infusion site
as programmed. The Occlusion Alarm Sensitivity is set to
High
by default.
Notice
a.
Next is highlighted.
Scroll to highlight
b.
Sensitivity.
Press
c.
Enter button.
Scroll to select
d.
High or Low.
Press
e.
Enter button.
Check that
32.
OCCLUSION
ALARM
sensitivity is
correct.
Notice your Sensitivity setting on the screen.
If setting is
Then do this
Correct
Notice Next is highlighted.
Press Enter button.
Incorrect
Repeat Step 31.
Set LOW INSULIN
33.
ALERT.
Notice
a.
Next is highlighted.
Scroll to highlight
b.
Low Alert (Number of units left in reservoir when pump
alarms).
Press
c.
Enter button.
Scroll to set low insulin
d.
alert
level.
Press
e.
Enter button.
Note:
The default is 20U. The range is 5U to 50U.