![Abbott FreeStyle Aviator Скачать руководство пользователя страница 54](http://html1.mh-extra.com/html/abbott/freestyle-aviator/freestyle-aviator_user-manual_3319185054.webp)
87
88
Steps
Do This
Troubleshooting Error
12.
Messages.
Error 1 (Er 1) Possible Causes:
Sample is too small.
•
Problem with the
•
FreeStyle Lite Test Strip.
Problem with the companion.
•
Very low blood glucose (less then 20 mg/dL [1.1 mmol/L]).
•
What You Should Do:
If you have symptoms such as weakness, sweating, nervousness, headache or
1.
confusion, follow your healthcare professional’s recommendation for treating
hypoglycemia.
Conduct a control solution test using a new
2.
FreeStyle Lite Test Strip. If the
results of the control solution test are within the appropriate range printed on
the side of your test strip vial, retest using blood and a new FreeStyle Lite Test
Strip.
If the retest using control solution does NOT work or the error persists, call
3.
Customer Care.
Error 2 (Er 2) Possible Causes:
Problem with the
•
FreeStyle Lite Test Strip.
Problem with the companion.
•
Very high blood glucose (above 500 mg/dL [27.8 mmol/L]).
•
HIGH control solution applied when temperature is too cold. (This applies only
•
to control solution that is labeled as HIGH).
What You Should Do:
If you have symptoms such as thirst, fatigue, excess urination or blurry
1.
vision, follow your healthcare professional’s recommendation for treating
hyperglycemia.
Conduct a control solution test using a new
2.
FreeStyle Lite Test Strip. If the
results of the control solution test are within the appropriate range printed on
your test strip vial, retest using blood and a new FreeStyle Lite Test Strip.
If the retest using control solution does
3.
NOT work or the error persists, call
Customer Care.
Steps
Do This
Troubleshooting Error
Messages (continued).
(Error 4 screenshot)
Error 3 (Er 3) Possible Causes:
Incorrect test procedure. For example, putting blood on the test strip before
•
inserting the test strip into the strip port or applying blood before the blood
drop and test strip symbols display.
Problem with the
•
FreeStyle Lite Test Strip.
Problem with the companion.
•
What You Should Do:
Be sure you see the blood drop and test strip symbols on the display screen
1.
before you apply blood or control solution.
Conduct a control solution test using a new FreeStyle Lite Test Strip. If the
2.
results of the control solution test are within the appropriate range printed on
your test strip vial, retest using blood and a new FreeStyle Lite Test Strip.
If the retest using control solution does NOT work or the error persists, call
3.
Customer Care.
Error 4 (Er 4) Possible Causes:
Problem with the
•
FreeStyle Lite Test Strip.
Problem with the companion.
•
What You Should Do:
Conduct a control solution test using a new
1.
FreeStyle Lite Test Strip. If the
results of the control solution test are within the appropriate range printed on
the side of your test strip vial, retest using blood and a new FreeStyle Lite Test
Strip.
If the retest using control solution does
2.
NOT work or the error persists, call
Customer Care.
Section 5: Testing Your Blood Glucose
5.2 Check Your Blood Glucose
Section 5: Testing Your Blood Glucose
5.2 Check Your Blood Glucose
FPO
DOC15736_Rev-A-Man.indd 87-88