What It Means
What To Do
2. The test
does not
start after
applying
the blood
drop.
•
Blood drop is too small
1. Add blood to the same sample area within 60
seconds, or repeat the test using a new test
strip and a larger blood drop.
•
Defective test strip
•
Sample applied after
meter turns off
1. Repeat the test using a new test strip. Wait for
and to appear before applying blood or
control solution.
2. If problem persists, contact Customer Care.
•
Defective meter
1. Contact Customer Care.
FreeStyle Control Solution
Purpose
FreeStyle control solution is a red liquid that contains a fixed amount of glucose.
The purpose of doing a control solution test is:
•
To practise testing
without
using your own blood
•
To confirm that the meter and test strips work together properly
•
When you are unsure of your blood glucose test results
Contact Customer Care for information on how to obtain control solution.
Control Solution Information
•
Use only FreeStyle control solution with the meter.
•
Replace the cap securely on the bottle immediately after use.
•
Control solution results are accurate only between 59° – 104° F (15° – 40° C).
•
Do
not
add water or other liquid to control solution.
•
There are three levels of control solution available (Low, Normal, High).
For information on how to obtain control solution, contact Customer Care.
IMPORTANT:
Control solution results should fall within the control
solution range printed on the test strip vial. Do
not
use this range
when testing blood glucose.