SECTION 6
i-STAT
Alinity — Quick Reference Guide
Art: 731848-01 Rev K Rev. Date: 02-Apr-2021
TROUBLESHOOTING
The i-STAT Alinity is programmed to perform quality checks
throughout the testing cycle.
The instrument has several methods of notifying operators
of failed quality checks.
1. Quality Check Failures
- Are displayed when the instrument identifies a problem
while running a cartridge or simulator
- There are 4 types of quality check failures:
1. Instrument
2. Cartridge
3. Sample
4. Software
- Screen displays the type of failure and instructions
for resolution
View Resolution
Cause
Excess blood added to cartridge.
When filling this cartridge, the blood
advanced past the level indicated by
the 'fill to' arrow.
Help
?
Options
Menu
Home
Cartridge Quality Check Failure
Cartridge Was Overfilled
Code: 37-01
View
Entered Info
Options
Menu
i-STAT CHEM8+
Pt: 654321
13OCT2025 08:33
View Cause
Cartridge Quality Check Failure
Resolution
Cartridge Was Overfilled
When filling a cartridge, use care to
advance blood to the level indicated by
the 'fill to' arrow.
Repeat testing with a freshly filled
cartridge.
Carefully observe the help provided
throughout the testing pathway.
If the same quality check failure displays,
contact the system administrator for
further instruction.
Code: 37-01
Help
?
Options
Menu
Home
View
Entered Info
Options
Menu
i-STAT CHEM8+
Pt: 654321
13OCT2025 08:33