SECTION 5
DISINFECTING
i-STAT Alinity Instrument
Disinfection is recommended between each patient. When instrument is
dedicated to a single patient, disinfect at least once a day. The disinfection
process must begin IMMEDIATELY after the cleaning procedure is
complete. Standard precautions should be taken whenever working with
blood or blood products.
Remove a wipe from the
container and squeeze out
any excess fluid before you
begin to wipe the surface.
Wipe all surfaces three times.
This includes front, back, sides,
top and bottom.
WIPE
WIPE
3X
3X
WIPE
WIPE
3X
3X
W
I
P
E
3
X
3
X
1
2
3
4
Wipe with gauze
until dry.
Allow all
surfaces to
remain wet
for 3 minutes.
SENSITIVE AREAS
Avoid forcing liquid into these areas:
i-STAT Alinity Instrument
Cartridge Port
10-Pin Connector under the camera
Gold contacts (2) on the outside of
the battery
A
B
C
C
B
A