15. When the procedure is complete, slowly remove the sheath. 

WARNING: Remove the sheath slowly to prevent an ingress of air.

Disposal

• The instructions for use are recyclable. Dispose of all packaging materials appropriately. 
• Dispose of delivery systems and accessories following standard solid biohazard waste procedures. 

Warranty

Abbott Medical warrants to buyer that, for a period equal to the validated shelf life of the product, this product shall meet the 
product specifications established by the manufacturer when used in accordance with the manufacturer's instructions for use 
and shall be free from defects in materials and workmanship. Abbott Medical's obligation under this warranty is limited to 
replacing or repairing at its option, at its factory, this product if returned within the warranty period to Abbott Medical and after 
confirmed to be defective by the manufacturer. 
EXCEPT AS EXPRESSLY PROVIDED IN THIS WARRANTY, ABBOTT MEDICAL DISCLAIMS ANY REPRESENTATION OR 
WARRANTY OF ANY KIND, EXPRESS OR IMPLIED, INCLUDING ANY WARRANTY AS TO MERCHANTABILITY OR 
FITNESS FOR A PARTICULAR PURPOSE. 
See the Terms and Conditions of Sale for further information.

State of California (USA) Only:

WARNING: This product can expose you to chemicals including ethylene oxide, which is known to the State of California to 
cause cancer and birth defects or other reproductive harm. For more information, go to www.P65Warnings.ca.gov.

Symbol Definitions

The following symbols may appear on the device packaging:

Symbol

Definition

Caution, consult accompanying documents

Manufacturer

Contains phthalate

EU authorized representative

Reference number

Product serial number

Batch Code

Use-by date 
(Use on or before the last day of the expiration month noted on the product packaging.)

Do not re-use

Sterilized using ethylene oxide

Unique device identification

Содержание Amplatzer Trevisio 9-ATV06F45

Страница 1: ...troduces an Amplatzer device into the sheath B Hemostasis valve with extension tube and stopcock Allows flushing of the delivery system and controls blood backflow C Sheath Provides a pathway through which an Amplatzer device is delivered D Dilator Eases penetration of tissue and minimizes vessel trauma E Delivery cable Attaches to the device to control its movement through the sheath F Plastic vi...

Страница 2: ... to avoid damaging tissue and vessels or interfering with previously implanted medical devices Use standard transcatheter techniques when using Amplatzer products Potential Adverse Events Potential adverse events that may occur during or after a procedure using this delivery system may include but are not limited to Device Compatibility The Amplatzer devices compatible with the Amplatzer Trevisio ...

Страница 3: ...or use 8 Turn the rotating luer on the dilator counterclockwise to unlock the dilator Slowly remove the dilator from the sheath WARNING Remove the dilator slowly to prevent an ingress of air 9 Allow blood backflow to purge any air from the sheath 10 Attach the hemostasis valve to the proximal end of the loader and flush with sterile saline CAUTION Use the hemostasis valve to impede blood backflow ...

Страница 4: ...tt Medical and after confirmed to be defective by the manufacturer EXCEPT AS EXPRESSLY PROVIDED IN THIS WARRANTY ABBOTT MEDICAL DISCLAIMS ANY REPRESENTATION OR WARRANTY OF ANY KIND EXPRESS OR IMPLIED INCLUDING ANY WARRANTY AS TO MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE See the Terms and Conditions of Sale for further information State of California USA Only WARNING This product can expo...

Страница 5: ...package is damaged Not made with natural rubber latex Inner diameter Outer diameter Length Usable length Recommended delivery sheath catheter dimensions Caution Federal law restricts this device to sale by or on the order of a physician Quantity Date of manufacture Intravascular Delivery System ONLY ...

Отзывы: