Directions for Use

Materials recommended for use with the delivery system

Exchange-length 0.035-inch guidewire

Procedure

CAUTION: When placing a device using an Amplatzer™ Trevisio™ Intravascular Delivery System, refer to the 
instructions for use provided with the device.

General instructions for the Amplatzer™ Trevisio™ Intravascular Delivery System are provided below.

 1.  Select the appropriate delivery system for use with the device being delivered. Refer to Table 1 to select a delivery 

system by Amplatzer™ device size, or refer to Table 2 to select the delivery system with a sheath that has the correct 
inner diameter as indicated in the device’s instructions for use.

2. Prepare the delivery system for use:

- Inspect the sterile pouch and verify that it is unopened and undamaged. Do not use the components if the sterile 

barrier has been compromised.

- Gently open the sterile pouch and inspect the components for damage. Do not use damaged or kinked components.
- Flush all components with sterile saline.
- Wipe the dilator and sheath with sterile gauze moistened with sterile saline to remove any foreign material.

3. Access the desired vessel.
4. Place the guidewire according to the device’s instructions for use. 
5. Insert the dilator into the sheath. You may encounter resistance as the dilator reaches the distal end of the sheath 

because the end of the sheath is tapered. 

6. Turn the rotating luer on the dilator clockwise to lock the components together.
7. Advance the dilator and sheath over the guidewire until the sheath is positioned according to the device’s instructions for 

use.

8. Turn the rotating luer on the dilator counterclockwise to unlock the dilator. Slowly remove the dilator from the sheath. 

WARNING: Remove the dilator slowly to prevent an ingress of air.

9. Allow blood backflow to purge any air from the sheath.

10. Attach the hemostasis valve to the proximal end of the loader and flush with sterile saline.

CAUTION: Use the hemostasis valve to impede blood backflow during the implant procedure.

11. Capture the device in the loader according to the device's instructions for use.
12. Attach the distal end of the loader to the proximal end of the sheath. Turn the rotating luer on the loader clockwise to lock 

the components together. Remember to close the stopcock on the hemostasis valve.

13. Advance the device until the black mark reaches the hemostasis valve, indicating the Amplatzer™ device is near the end 

of the delivery system.
WARNING: When the black mark reaches the hemostasis valve, some devices and device sizes may be closer to the 
end of the delivery system due to differences in device, sheath, and loader lengths. Advancing the device without 
fluoroscopy may result in patient injury.

14. Position, deploy, and detach the device according the device’s instructions for use.

Table 2. Delivery System Dimensions

Delivery system 

(sheath) size

Inner diameter of 

sheath

Outer diameter of 

sheath

Model Number/Delivery 

System Size (mm)

6 Fr

2.11 mm (0.08 in)

2.79 mm (0.11 in)

9-ATV06F45/60

7 Fr

2.44 mm (0.10 in)

3.18 mm (0.13 in)

9-ATV07F45/60

7 Fr

2.44 mm (0.10 in)

3.18 mm (0.13 in)

9-ATV07F45/80

8 Fr

2.69 mm (0.11 in)

3.45 mm (0.14 in)

9-ATV08F45/60

8 Fr

2.69 mm (0.11 in)

3.45 mm (0.14 in)

9-ATV08F45/80

9 Fr

3.00 mm (0.12 in)

3.81 mm (0.15 in)

9-ATV09F45/80

10 Fr

3.30 mm (0.13 in)

4.14 mm (0.16 in)

9-ATV10F45/80

12 Fr

3.99 mm (0.16 in)

4.80 mm (0.19 in)

9-ATV12F45/80

Содержание Amplatzer Trevisio 9-ATV06F45

Страница 1: ...troduces an Amplatzer device into the sheath B Hemostasis valve with extension tube and stopcock Allows flushing of the delivery system and controls blood backflow C Sheath Provides a pathway through which an Amplatzer device is delivered D Dilator Eases penetration of tissue and minimizes vessel trauma E Delivery cable Attaches to the device to control its movement through the sheath F Plastic vi...

Страница 2: ... to avoid damaging tissue and vessels or interfering with previously implanted medical devices Use standard transcatheter techniques when using Amplatzer products Potential Adverse Events Potential adverse events that may occur during or after a procedure using this delivery system may include but are not limited to Device Compatibility The Amplatzer devices compatible with the Amplatzer Trevisio ...

Страница 3: ...or use 8 Turn the rotating luer on the dilator counterclockwise to unlock the dilator Slowly remove the dilator from the sheath WARNING Remove the dilator slowly to prevent an ingress of air 9 Allow blood backflow to purge any air from the sheath 10 Attach the hemostasis valve to the proximal end of the loader and flush with sterile saline CAUTION Use the hemostasis valve to impede blood backflow ...

Страница 4: ...tt Medical and after confirmed to be defective by the manufacturer EXCEPT AS EXPRESSLY PROVIDED IN THIS WARRANTY ABBOTT MEDICAL DISCLAIMS ANY REPRESENTATION OR WARRANTY OF ANY KIND EXPRESS OR IMPLIED INCLUDING ANY WARRANTY AS TO MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE See the Terms and Conditions of Sale for further information State of California USA Only WARNING This product can expo...

Страница 5: ...package is damaged Not made with natural rubber latex Inner diameter Outer diameter Length Usable length Recommended delivery sheath catheter dimensions Caution Federal law restricts this device to sale by or on the order of a physician Quantity Date of manufacture Intravascular Delivery System ONLY ...

Отзывы: