1. If any of the lead or extension contacts came in contact with body fluid or saline, thoroughly clean the
1.
contacts with sterile deionized water or sterile water for irrigation and dry them completely.
CAUTION: Do not bend the lead sharply or it may be damaged.
NOTE: The EPG header comes with open ports and attached port plugs. Do not remove the port plugs
from the header. If only using one lead or extension, use the lower port labeled 1
‑
8 and insert a port
plug in the unused port.
2. Push and hold the button on the EPG header to open the EPG header ports.
2.
3. Using clean gloves, carefully slide the lead or extension into the EPG header until you reach a hard stop
3.
and all the contact bands are fully inside the connector assembly and hidden from view.
25