Do not put heavy or sharp objects on the EPG.
▪
▪
Keep the EPG away from food and beverages, and avoid getting the EPG wet.
▪
▪
If the EPG does get wet, do not use it. Contact Technical Support for help.
▪
Cleaning the EPG
▪
Using two PDI Sani-Cloths, wipe the EPG for 2 minutes front and back, then allow to air dry. Once the
▪
EPG is dry, attach the header cap.
NOTE: Clean the EPG before attaching the header cap.
▪
Clean the EPG by thoroughly wiping off its outer surface using the provided cleaning cloths. Do not wipe
▪
over the stimulator port. Do not submerge the EPG in fluid or use a cloth that is soaking wet.
▪
Never use ammonia, acidic, or alkaline cleaners or organic chemicals such as paint thinner, acetone,
▪
propyl alcohol, or kerosene. They may damage surface finishes of the EPG.
Storing and Transporting the EPG
▪
Store the EPG and batteries in a cool, dry place. Do not store the EPG or batteries in your car or garage.
▪
▪
Transport the EPG in a carrying case.
▪
▪
If you are carrying the EPG in a briefcase or other carrying case, make sure no objects in the case are
▪
pressing on the device.
▪
Store paired EPGs and patient controllers together to reduce the number of times you need to set up
▪
communication.