background image

10

Normal Operation

Pt/Pt 13% Rh Thermocouple

NiCr/NiAl

Thermocouple

Non-Carburizing
Atmospheres only

400

600

800

1000

1200

1250

1400

150

100

50

Kilocalories

Probe Output
(mV)

1500

500

1000

Degrees Celsius

Electronic Conduction

Errors

Probe

Life

Reduced

High

Resistance

752

1400

1700

2552

Degrees Fahrenheit

1112

2000

2282

5

OPERATION

6

MAINTENANCE

5.1

Principles of Operation

The ZGP2 probe contains a high temperature oxygen
concentration cell using zirconium oxide as a solid electrolyte.
Inner and outer electrodes are attached to the cell which is
specific to oxygen.

Air is supplied to the inner reference electrode to provide a
constant partial pressure of oxygen while the process gas to
be measured is in contact with the outer electrode.

A voltage is generated across the electrodes which is a
function of the ratio of the oxygen partial pressures at the two
electrodes. This voltage output represents ‘oxygen potential’
and can be used to control the atmosphere of annealing
processes etc.

For the measurement of ‘oxygen concentration’ it is necessary
to correct for absolute temperature of the probe. The
temperature term in the Nernst equation, on which the voltage
output depends, illustrates this point:

E = 0.0496 T (log

10

 P

0

/P

1

) + C mV

Where:

T

= Absolute Temperature (

°

K)

P

0

= Partial Pressure Reference Gas

P

1

= Partial Pressure Sample Gas

C = Cell Constant

For  ‘oxygen potential’ measurement the following two
expressions relate the oxygen probe output (E mV) to oxygen
potential (

µ

)

E = (10.84

µ

 + 40) where 

µ

 is in kilocalories

or

E = –(2.591

µ

 + 40) where 

µ

 is in kilojoules

5.2

Range of Operation – Fig. 5.1

The internal resistance of the cell, i.e the resistance of the
electrolyte between the electrodes, decreases approximately
exponentially with increasing temperature and for this reason
it is recommended that the probe is used at temperatures
greater than 600

°

C (1112

°

F).

The maximum operating temperature is limited by two factors:

a) The onset of electronic conduction through the electrolyte

which reduces the measured output below its theoretical
value.

b) Evaporation of the outer electrode.

Electronic conduction is a function of both temperature and
oxygen partial pressure and, for stabilized zirconia, it occurs at
low levels of oxygen and high temperature. For the above
reasons it is recommended that the probe is used at
temperatures between 600

°

C (1112

°

F) and 1250

°

C (2282

°

F) –

see Fig. 5.1. [900

°

C (1652

°

F) max. for Type K thermocouple].

Caution. 

No maintenance is necessary, or

possible, on the probe and any attempt to dismantle it
could cause irreparable damage. The probe can be tested
for accuracy whilst connected to its measuring system but
without removal from its operating position.

It is recommended that at least one spare probe is held for
replacement or comparison purposes.

Fig. 5.1 Range of Operation

Содержание ZGP2 Series

Страница 1: ...AnalyzeIT High Temperature Zirconia Oxygen Probe ZGP2 Series Operating Instructions IM ZGP2_8 ...

Страница 2: ...l conditions result in degraded process system performance leading to personal injury or death Therefore comply fully with all Warning and Caution notices Information in this manual is intended only to assist our customers in the efficient operation of our equipment Use of this manual for any other purpose is specifically prohibited and its contents are not to be reproduced in full or part without...

Страница 3: ...librium shift common to other systems in which gas samples are cooled before measurement and has a fast response to changes in atmosphere enabling rapid corrective action to be taken when necessary The probe comprises a ceramic detector cell housed in a protective sheath A thermocouple is fitted within the probe to enabletheprocesstemperaturetobemonitoredorforautomatic temperature compensation to ...

Страница 4: ...h is fitted adjust the diameter of the wadding at the end of the probe to allow a snug fit To reduce its diameter the wadding may be compressed slightly To increase its diameter apply slight end pressure on the wire bonding at both ends of the wadding Mould the wadding by hand to ensure a light push fit into the sheath 2 2 Checking the Code Number Ensure that the correct equipment is being install...

Страница 5: ...3 Wadding Filter Washer Vent Holes 1 to 1 5 mm 1 3 2 5 6 7 8 8 3 4 5 8 2 2 PREPARATION Fig 2 1 Assembling the Probe ...

Страница 6: ...urement is required i e O2 ppmO2 or O2 partial pressure temperature compensation of the probe output is usually necessary If the process temperature is constant or if variations in operating temperature do not produce unacceptable errors as in some flue gas measurement it is possible to measure O2 concentration without temperature compensation using a system similar to that for oxygen potential Sy...

Страница 7: ...uple mV Output from Zirconia Cell Thermocouple Output from Probe Z MT Oxygen Analyzer Alarm Control or Retransmission Outputs Alarm Control or Retransmission Outputs Alarm Control Retransmission or Serial Data Output Power Supply Remote Fuel Selector dual fuel versions only Carbon Monoxide Monitor 3 INSTALLATION Fig 3 2 Oxygen Concentration Measurement using Basic Oxygen Analyzer Fig 3 3 Oxygen Co...

Страница 8: ... subsequent removal Dimensions for the probe are shown in Figs 3 4 to 3 6 A clearance of at least 25mm in excess of the overall probe length is necessary for installation or removal procedures Caution The probe MUST NOT be sited a where it is subjected to mechanical or thermal shock b in the presence of aggressive components e g molten slags molten silicates metals and vapours of lead zinc silicon...

Страница 9: ...ustomer M16 Cable Entries x2 Reference Air Line Entry Mounting Flange Probe Outer Sheath Test Gas All dimensions in mm inches 3 INSTALLATION 3 2 Siting Figs 3 4 to 3 6 Fig 3 5 Overall Dimensions Twin Gland C95 Head Fig 3 6 Overall Dimensions 1 in NPT Adaptor 69 8 2 74 View A A 15 0 59 Twin Gland Head Shown for Information Only All dimensions in mm inches ...

Страница 10: ...e of the probe including any stand off which may be used is thermally insulated to minimize condensation within the probe see Fig 3 7 Condensation within the probe sheath particularly in outdoor installations can be sufficient to allow water to come into contact with the hot ceramic zirconia tube leading to thermal shock and failure of the sensor This problem is most likely to occur on such instal...

Страница 11: ...able for use with Pt Pt 13 Rh thermocouples to BS4937 sheathing to BS6746 part no E37 4 3 Electrical Connections Fig 4 2 Make connections as shown in Fig 4 2 and on the label inside the probe head cover The connecting cable must enter the probe head via the bush provided standard head or suitable cable glands twin gland C95 head 4 4 Reference Air Connections Fig 4 2 The reference air connector is ...

Страница 12: ...ference Gas P1 Partial Pressure Sample Gas C Cell Constant For oxygen potential measurement the following two expressions relate the oxygen probe output E mV to oxygen potential µ E 10 84µ 40 where µ is in kilocalories or E 2 591µ 40 where µ is in kilojoules 5 2 Range of Operation Fig 5 1 The internal resistance of the cell i e the resistance of the electrolyte between the electrodes decreases app...

Страница 13: ...nnections b Remove the screwed plug from the test gas connector see Fig 4 2A c Fit 1 4in i d x 3 8in o d plastic or similar tubing and supply a test gas of known oxygen concentration to the probe at a steady flow rate of 800 to 1000ml min 1 7 to 2 1ft3 hr Allow at least 5 minutes for the system to stabilize before making a measurement d Check that the oxygen concentration measured by the probe sys...

Страница 14: ...he limits are given in Table 8 1 for 1 electronic conduction but in practice it may be possible to measure even lower oxygen potential levels without introducing significant errors Electrical Data Cell Constant 0V 2mV Probe accuracy Typically better than 5 of reading System accuracy 2 of test gas reading when calibrated against a certified test gas Cell output impedance Typically 100kΩ 700 C 1292 ...

Страница 15: ...umentation pH conductivity and dissolved oxygen transmitters and sensors ammonia nitrate phosphate silica sodium chloride fluoride dissolved oxygen and hydrazine analyzers Zirconia oxygen analyzers katharometers hydrogen purity and purge gas monitors thermal conductivity Customer Support We provide a comprehensive after sales service via a Worldwide Service Organization Contact one of the followin...

Страница 16: ... Lockheed Way Carson City NV 89706 USA Tel 1 775 883 4366 Fax 1 775 883 4373 ABB has Sales Customer Support expertise in over 100 countries worldwide www abb com The Company s policy is one of continuous product improvement and the right is reserved to modify the information contained herein without notice Printed in UK 03 04 ABB 2004 ...

Отзывы: