GB6
Temperature Sensor TSHD (SensyTemp Exd)
42/10-55 XU
– Verify used cable for suitability (type, actual outside diameter, etc).
– Tighten cable gland until the cable is firmly fixed and sealed by the sealing ring.
– Protect cable against mechanical loads (tension, torsion, etc) with suitable installation
measures.
– Close openings, which are not needed with certified plugs.
Grounding terminals
Additional grounding connections are not necessary in case of reliable electrical contact to
grounded conductive tubing systems or other grounded conductive supporting structures (e.g.
via connection threads or flange screws).
Electrical components
– Assemble electrical components with assembling sets (if needed).
– Connect electrical components according to the relevant operating instructions.
Cover
– Screw on cover manually until O seal is compressed (check threads for damages before)
and secure cover with the locking screw.
– Do not use any wrenches to tighten the cover (hexagons and recesses are only opening
aids).
4.2
Assembling of field housings AGLF, AGLFH, AGLFD
Installation
Install field housings permanently to fixed walls, supports or pipes.
Cable glands
– Use only approved cable glands of the proper protection class.
– Observe certification and installation instructions of the cable glands.
– Screw in cable glands with appropriate thread securer (e.g. Loctite 273, not necessary for
NPT threads, check threads for damages before).
– Verify used cable for suitability (type, actual outside diameter, etc.).
– Tighten cable gland until the cable is firmly fixed and sealed by the sealing ring.
– Protect cable against mechanical loads (tension, torsion, etc.) with suitable installation
measures.
– Close openings, which are not needed with certified plugs.
Grounding terminals
Additional grounding connections are not necessary in case of reliable electrical contact to
grounded conductive tubing systems or other grounded conductive supporting structures (e.g.
via metallic mounting plate).
Electrical components
– Assemble electrical components with assembling sets (if needed).
– Connect electrical components according to the relevant operating instructions.
Cover
– Screw on cover manually until O seal is compressed (check threads for damages before)
and secure cover with the locking screw.
– Do not use any wrenches to tighten the cover (hexagons and recesses are only opening
aids).
Содержание TSHD
Страница 10: ...D8 Temperaturfühler TSHD SensyTemp Exd 42 10 55 XU 7 Anlage 1 EG Baumusterprüfbescheinigung Seite1 ...
Страница 11: ...42 10 55 XU Temperaturfühler TSHD SensyTemp Exd D9 8 Anlage 2 Konformitätserklärung ...
Страница 19: ...GB8 Temperature Sensor TSHD SensyTemp Exd 42 10 55 XU 7 Enclosure 1 EC Type Examination page 1 ...
Страница 20: ...42 10 55 XU Temperature Sensor TSHD SensyTemp Exd GB9 8 Enclosure 2 Declaration of conformity ...