* These screens and tests apply to the Liquid Orifice Tube
Measurement application that is running on G3, XFC and XRC.
5.2 Verifying
Configuration
The following test requires that a liquid measurement tube be enabled and configured
correctly. Holding registers also need to be enabled within PCCU.
1)
Within PCCU, click on Holding Registers. Upon opening the feature, click the
Capacity tab. Verify that Descriptors are enabled for ‘floats’.
2)
Next, within the Holding Registers feature, click on the Float tab.
3)
Within the Float tab, locate Float 0. The user will notice that the
app/array/register is 9.0.0 (see Figure 11). In the Description field, type
DP
.
Move to the corresponding Value field, and enter in a value that the user wants
used for the differential pressure for this test.
Figure 11 Float Tab
4)
Within the same tab, locate the app/array/register 9.0.1. In the Description field,
type SP. Move to the corresponding Value field, and enter in a value that the
user wants used for the static pressure for this test.
5)
Within the same tab, locate the app/array/register 9.0.2. In the Description field,
type TF. Move to the corresponding Value field, and enter in a value that the
user wants used for the temperature for this test.
6)
When all the information has been entered, click the Send button.
7)
The user will need to move back to the Local Connect screen. The values that
were entered into the holding registers should appear within the corresponding
boxes in the Setup tab (See Figure 12).