Page 10
2103956 Rev. AA
Figure 8 Calibration Setup Screen
The static pressure, differential pressure and temperature registers must be set to I/O
or XMV registers to allow for calibration.
4.1 Calibration
Checks
Tab
The Calibration Checks tab (see Figure 9) allows the current measured values to be
tested and logged as events when they are compared to a known benchmark test.
Figure 9 Calibration Checks Tab
Clicking the Enter Pressure Check Value button will display the Enter Pressure dialog
box. Within the box, the user can enter a corresponding value. When the measured
value stabilizes, clicking the OK button will log the device’s measured value, and the
customer entered value. This value is then displayed in the Checks grid.