Downloading and Installing the

Operating System Software – Figs. 4.1 to 4.5

14) Isolate the instrument from the power supply.

15) Remove the instrument from its case – see Fig. 4.1.

16) Push the red switch rearwards  – see Fig. 4.1.

17) Refit the instrument to the case – see Fig. 4.1.

18) Restore the power supply to the instrument.

19) Follow the on-screen instructions and perform

touchscreen calibration.

Caution. Do not use sharp objects such as

screwdrivers, pen nibs etc. to operate the touch screen.

Note. This final calibration must be very accurate as

it is the calibration which will be saved by the recorder and
used from this point forward to operate the instrument.

Содержание SmartMedia SM2000

Страница 1: ...alidating the instrument s warranty 2 Tools Required Medium flat bladed screwdriver No 1 Pozidriv screwdriver Fine tipped soldering iron 1 Introduction This information sheet describes the procedure f...

Страница 2: ...ng screw 1 2 3 Information Refitting is the reverse of removal Fig 3 1 Removing the Instrument from its Case 3 Remove the chassis top cover and PCB locking bar see Fig 3 2 3 Fitting Media Board Sub As...

Страница 3: ...w the main board from the chassis 3 Disconnect the touchscreen connector from the main board 2 Slot D 1 Connection locking bar Note To disconnect the display flexi circuit it may be necessary to remov...

Страница 4: ...out by a Company Approved Service Engineer 3 Fitting Media Board Sub Assembly Figs 3 1 to 3 7 5 Refer to Fig 3 4 and check the part number of the Main Board ve leg below C14 Reverse of Main Board Show...

Страница 5: ...is already fitted to the instrument Fig 3 5 Removing the Media Board Sub Assembly Information Refitting is the reverse of removal 1 Check that PCB link LK1 is fitted and ensure it is in the correct po...

Страница 6: ...5 Caution When positioning the chassis top cover take care not to damage the EMC contacts Fig 3 7 Refitting the PCB Locking Bar and Chassis Top Cover 3 Fitting Media Board Sub Assembly Figs 3 1 to 3 7...

Страница 7: ...e instrument Operating System requires upgrading to 2002 issue 8 0 or later Application Code and 1002 issue 6 0 or later System Code see Fig 1 1 for details of checking the Code versions 1 Isolate the...

Страница 8: ...Internal File View button 3 Insert a memory card containing a folder named SM2000 2 4 Downloading and Installing the Operating System Software Figs 4 1 to 4 5 4 Refit the instrument to the case see Fi...

Страница 9: ...ect the PC Configurator Cable Part No C100 0051 to the RS232 Serial Port of the PC 3 Fig 4 3 Connecting the PC Configurator Cable 4 Downloading and Installing the Operating System Software Figs 4 1 to...

Страница 10: ...rt Recorders Fig 4 5 Creating the Binary Partition 11 Restore the power supply to the instrument and start the download of the new software by clicking on the Program button on the SM2000 Serial Downl...

Страница 11: ...fit the instrument to the case see Fig 4 1 18 Restore the power supply to the instrument 19 Follow the on screen instructions and perform touchscreen calibration Caution Do not use sharp objects such...

Страница 12: ...ntained herein without notice Printed in UK 02 02 ABB 2002 ABB Ltd Howard Road St Neots Cambridgeshire PE19 8EU UK Tel 44 0 1480 475 321 Fax 44 0 1480 217 948 ABB Inc 125 E County Line Road Warminster...

Отзывы: