Communication
28 Sensyflow
FMT700-P
42/14-42-EN
Pos: 15.13 /Kommunikation/Durchfluss/FMT700-P/Allgemeine Informationen @ 33\mod_1253169393484_3101.doc @ 301811 @ 4
6.1.1.5 General
information
Enter the command
XTT
CR
to access general information about the interface. When a monitor
is connected, the following text appears:
CLS, IV General information
Vers. X.XX
Progr. Sensyflow FMT700-P
Digital interfaces
15-bit data ready
12-bit digital out / optically decoupled
Serial interface
RS 232 / V 24
Serial interface command list
<XB>
Analog interface
0 ... 10 V or 0 ... 20 mA / 4 ... 20 mA
Digital coding of flowmeter sensor
size
E.g., DN 25 Sensor 1
1100
1100
Measuring rate for display and
analog output changed via BCD
switch
Measuring rate for display
0100 ... 9999 measuring cycles
Measuring rate for analog out
0001 ... 9999 measuring cycles
Fast mode
Display not used; each measured value is sent to the
analog output
Temperature option
Measuring range -20 ... 100 °C
2-probe operation
Switchover between 2 sensors
Sensor / Probe = flowmeter sensor
Pos: 15.14 /Kommunikation/Durchfluss/FMT700-P/Fehlertabelle @ 33\mod_1253169451593_3101.doc @ 301835 @ 4
6.1.1.6 Error
table
Enter the command
XF
CR
to access the error table. When a monitor is connected, the following
text appears:
CLS
"Sensyflow FMT700-P" error table
Error
Description
999999 Measured
value
≥
measuring range (normal operation)
E02 No
probe
connected
E03 Sum
register
overflow
E04
A measured value > probe's URL
E11
No transmitter connected
E12
Pt100 sensor break
E13 ...
E14 ...
E15 ...
E16 ...
E17 ...
E18 ...
E19 ...
E20 ...
Sensor / Probe / Transmitter = flowmeter sensor
URL = Upper range limit