![ABB Sensyflow FMT700-P Скачать руководство пользователя страница 15](http://html.mh-extra.com/html/abb/sensyflow-fmt700-p/sensyflow-fmt700-p_operating-instruction_4507015.webp)
Electrical
connections
42/14-42-EN Sensyflow
FMT700-P
15
Nominal diameter coding
Pin 5
Pin 18
Pin 6
Pin 19
Hex
No measuring tube
0 0 0 0 00
DN
25
1 1 0 0 30
DN
50
1 0 1 0 50
DN
80
1 1 1 0 70
DN
100
1 0 0 1 90
DN
150
1 1 0 1 B0
DN
200
1 0 1 1 D0
Special
1 1 1 1 F0
Pin 4
Pin 17
Flowmeter sensor 1
1 1
Flowmeter sensor 2
1 0
Medium temperature flowmeter sensor 1
0 1
Medium temperature flowmeter sensor 2
0 0